Importance In patients with severe aortic valve stenosis at intermediate surgical risk, transcatheter aortic valve replacement (TAVR) with a self-expanding supra-annular valve was noninferior to surgery for all-cause mortality or disabling stroke at 2 years. Comparisons of longer-term clinical and hemodynamic outcomes in these patients are limited. Objective To report prespecified secondary 5-year outcomes from the Symptomatic Aortic Stenosis in Intermediate Risk Subjects Who Need Aortic Valve Replacement (SURTAVI) randomized clinical trial. Design, Setting, and Participants SURTAVI is a prospective randomized, unblinded clinical trial. Randomization was stratified by investigational site and need for revascularization determined by the local heart teams. Patients with severe aortic valve stenosis deemed to be at intermediate risk of 30-day surgical mortality were enrolled at 87 centers from June 19, 2012, to June 30, 2016, in Europe and North America. Analysis took place between August and October 2021. Intervention Patients were randomized to TAVR with a self-expanding, supra-annular transcatheter or a surgical bioprosthesis. Main Outcomes and Measures The prespecified secondary end points of death or disabling stroke and other adverse events and hemodynamic findings at 5 years. An independent clinical event committee adjudicated all serious adverse events and an independent echocardiographic core laboratory evaluated all echocardiograms at 5 years. Results A total of 1660 individuals underwent an attempted TAVR (n = 864) or surgical (n = 796) procedure. The mean (SD) age was 79.8 (6.2) years, 724 (43.6%) were female, and the mean (SD) Society of Thoracic Surgery Predicted Risk of Mortality score was 4.5% (1.6%). At 5 years, the rates of death or disabling stroke were similar (TAVR, 31.3% vs surgery, 30.8%; hazard ratio, 1.02 [95% CI, 0.85-1.22]; P = .85). Transprosthetic gradients remained lower (mean [SD], 8.6 [5.5] mm Hg vs 11.2 [6.0] mm Hg; P < .001) and aortic valve areas were higher (mean [SD], 2.2 [0.7] cm 2 vs 1.8 [0.6] cm 2 ; P < .001) with TAVR vs surgery. More patients had moderate/severe paravalvular leak with TAVR than surgery (11 [3.0%] vs 2 [0.7%]; risk difference, 2.37% [95% CI, 0.17%- 4.85%]; P = .05). New pacemaker implantation rates were higher for TAVR than surgery at 5 years (289 [39.1%] vs 94 [15.1%]; hazard ratio, 3.30 [95% CI, 2.61-4.17]; log-rank P < .001), as were valve reintervention rates (27 [3.5%] vs 11 [1.9%]; hazard ratio, 2.21 [95% CI, 1.10-4.45]; log-rank P = .02), although between 2 and 5 years only 6 patients who underwent TAVR and 7 who underwent surgery required a reintervention. Conclusions and Relevance Among intermediate-risk patients with symptomatic severe aortic stenosis, major clinical outcomes at 5 years were similar for TAVR and surgery. TAVR was associated with superior hemodynamic valve performance but also with more paravalvular leak and valve reinterventions.
The transseptal puncture experience: Safety insights from FDA MAUDE database
AbstractBackgroundThere is a paucity of literature on safety and efficacy of various transseptal puncture (TSP) needles.ObjectivesTo assess the reported mechanisms of failure, complications, and outcomes among the most frequently used transseptal needles in the United States.MethodsWe queried the Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database between January 2011 and January 2021 for reports on the most commonly used transseptal needles: NRG (Baylis Medical, Montreal, Canada), and BRK (St. Jude, Saint Paul, MN)]. The primary outcome was the mechanism of failure. Secondary outcomes included clinical consequences of device failure.ResultsThe final analysis included 306 reports of failure/complication with TSP needles (NRG n = 70, BRK n = 236). The most commonly reported mode of failure was detachment of the needle component (i.e., clip, hub, stopcock, shaft, spring, or needle tip) (14.7% overall; 17.8% BRK; and 4.3% NRG). Among these reports, cardiac perforation was the most common complication (69.9% overall; 69.1% for BRK; and 72.9% for NRG). Pericardiocentesis was the second most commonly reported complication (45.1% overall; 48.3% for BRK; and 34.3% for NRG). The procedure was successfully completed in 33.3% of all cases (36.4% for BRK and 22.9% for NRG), while surgical conversion was needed in (13.4% overall; 14% for BRK and 11.4% for NRG) of the reports. Death occurred in 3.9% of all cases overall (3.4% for BRK and 5.7% for NRG).ConclusionsNeedle detachment was the most common mode of failure, and cardiac perforation was the most common complication reported with TSP needles. Future efforts should focus on innovative TSP needle design, best practice guidelines, including role of imaging guidance, and increased TSP training.
Device-Related Thrombus After Left Atrial Appendage Closure
Background: In patients with atrial fibrillation, left atrial appendage closure with the Watchman device prevents thromboembolism from the left atrial appendage; however, thrombus may form on the left atrial face of the device, and then potentially embolize. Herein, we studied the incidence, predictors, and clinical outcome of device-related thrombus (DRT) using a large series of clinical trial cohorts of patients undergoing Watchman implantation. Methods: We studied the device arms of 4 prospective Food and Drug Administration trials: PROTECT-AF (Watchman Left Atrial Appendage System for Embolic Protection in Patients With Atrial Fibrillation) (n=463); PREVAIL (Evaluation of the Watchman LAA Closure Device in Patients With Atrial Fibrillation Versus Long Term Warfarin Therapy) (n=269); CAP (Continued Access to PROTECT AF registry) (n=566); and CAP2 (Continued Access to PREVAIL registry) (n=578). Surveillance transesophageal echocardiographs were performed at 45 days and 12 months in all patients, and also at 6 months in the randomized control trials. We assessed both the incidence of DRT during these transesophageal echocardiographs (and other unscheduled transesophageal echocardiographs), and clinical outcomes of postprocedure stroke or systemic embolism (SSE) and adjusted for CHA 2 DS 2 -VASC and HAS-BLED scores. Results: Of 1739 patients who received an implant (7159 patient-years follow-up; CHA 2 DS 2 -VASc=4.0), DRT was seen in 65 patients (3.74%). The rates of SSE with and without DRT were 7.46 and 1.78 per 100 patient-years (adjusted rate ratio, 3.55; 95% confidence interval [CI], 2.18–5.79; P <0.001), and ischemic SSE rates were 6.28 and 1.65 per 100 patient-years (adjusted rate ratio, 3.22; 95% CI, 1.90–5.45, P <0.001). On multivariable modeling analysis, the predictors of DRT were as follows: history of transient ischemic attack or stroke (odds ratio [OR], 2.31; 95% CI, 1.26–4.25; P =0.007), permanent atrial fibrillation (OR, 2.24; 95% CI, 1.19–4.20; P =0.012); vascular disease (OR, 2.06; 95% CI, 1.08–3.91; P =0.028); left atrial appendage diameter (OR, 1.06 per mm increase; 95% CI, 1.01–1.12; P =0.019); left ventricular ejection fraction (OR, 0.96 per 1% increase; 95% CI, 0.94–0.99; P =0.009). DRT and SSE both occurred in 17 of 65 patients (26.2%). Of the 19 SSE events in these patients with DRT, 9 of 19 (47.4%) and 12 of 19 (63.2%) occurred within 1 and 6 months of DRT detection. Conversely, after left atrial appendage closure, most SSEs (123/142, 86.62%) occurred in patients without DRT. Conclusions: After left atrial appendage closure with Watchman, DRT (≈3.7%) is not frequent but, when present, is associated with a higher rate of stroke and systemic embolism.
Outcomes in the Randomized CoreValve US Pivotal High Risk Trial in Patients With a Society of Thoracic Surgeons Risk Score of 7% or Less
Abstract Objectives To evaluate the periprocedural characteristics and outcomes of patients supported with Impella 2.5 prior to percutaneous coronary intervention (pre‐PCI) versus those who received it after PCI (post‐PCI) in the setting of cardiogenic shock (CS) complicating an acute myocardial infarction (AMI). Background Early mechanical circulatory support may improve outcome in the setting of CS complicating an AMI. However, the optimal timing to initiate hemodynamic support has not been well characterized. Methods Data from 154 consecutive patients who underwent PCI and Impella 2.5 support from 38 US hospitals participating in the USpella Registry were included in our study. The primary end‐point was survival to discharge. Secondary end‐points included assessment of patients' hemodynamics and in‐hospital complications. A multivariate regression model was used to identify independent predictors for mortality. Results Both groups were comparable except for diabetes (P = 0.02), peripheral vascular disease (P = 0.008), chronic obstructive pulmonary disease (P = 0.05), and prior stroke (P = 0.04), all of which were more prevalent in the pre‐PCI group. Patients in the pre‐PCI group had more lesions (P = 0.006) and vessels (P = 0.01) treated. These patients had also significantly better survival to discharge compared to patients in the post‐PCI group (65.1% vs.40.7%, P = 0.003). Survival remained favorable for the pre‐PCI group after adjusting for potential confounding variables. Initiation of support prior to PCI with Impella 2.5 was an independent predictor of in‐hospital survival (Odds ratio 0.37, 95% confidence interval: 0.17–0.79, P = 0.01) in multivariate analysis. The incidence of in‐hospital complications included in the secondary end‐point was similar between the 2 groups. Conclusions The results of our study suggest that early initiation of hemodynamic support prior to PCI with Impella 2.5 is associated with more complete revascularization and improved survival in the setting of refractory CS complicating an AMI. (J Interven Cardiol 2014;27:1–11)