Abstract The semi‐adiabatic localization by adiabatic selective refocusing (sLASER) sequence provides single‐shot full intensity signal with clean localization and minimal chemical shift displacement error and was recommended by the international MRS Consensus Group as the preferred localization sequence at high‐ and ultra‐high fields. Across‐vendor standardization of the sLASER sequence at 3 tesla has been challenging due to the B 1 requirements of the adiabatic inversion pulses and maximum B 1 limitations on some platforms. The aims of this study were to design a short‐echo sLASER sequence that can be executed within a B 1 limit of 15 μT by taking advantage of gradient‐modulated RF pulses, to implement it on three major platforms and to evaluate the between‐vendor reproducibility of its perfomance with phantoms and in vivo. In addition, voxel‐based first and second order B 0 shimming and voxel‐based B 1 adjustments of RF pulses were implemented on all platforms. Amongst the gradient‐modulated pulses considered (GOIA, FOCI and BASSI), GOIA‐WURST was identified as the optimal refocusing pulse that provides good voxel selection within a maximum B 1 of 15 μT based on localization efficiency, contamination error and ripple artifacts of the inversion profile. An sLASER sequence (30 ms echo time) that incorporates VAPOR water suppression and 3D outer volume suppression was implemented with identical parameters (RF pulse type and duration, spoiler gradients and inter‐pulse delays) on GE, Philips and Siemens and generated identical spectra on the GE ‘Braino’ phantom between vendors. High‐quality spectra were consistently obtained in multiple regions (cerebellar white matter, hippocampus, pons, posterior cingulate cortex and putamen) in the human brain across vendors (5 subjects scanned per vendor per region; mean signal‐to‐noise ratio > 33; mean water linewidth between 6.5 Hz to 11.4 Hz). The harmonized sLASER protocol is expected to produce high reproducibility of MRS across sites thereby allowing large multi‐site studies with clinical cohorts.
Methodological consensus on clinical proton MRS of the brain: Review and recommendations
Proton MRS (1H MRS) provides noninvasive, quantitative metabolite profiles of tissue and has been shown to aid the clinical management of several brain diseases. Although most modern clinical MR scanners support MRS capabilities, routine use is largely restricted to specialized centers with good access to MR research support. Widespread adoption has been slow for several reasons, and technical challenges toward obtaining reliable good‐quality results have been identified as a contributing factor. Considerable progress has been made by the research community to address many of these challenges, and in this paper a consensus is presented on deficiencies in widely available MRS methodology and validated improvements that are currently in routine use at several clinical research institutions. In particular, the localization error for the PRESS localization sequence was found to be unacceptably high at 3 T, and use of the semi‐adiabatic localization by adiabatic selective refocusing sequence is a recommended solution. Incorporation of simulated metabolite basis sets into analysis routines is recommended for reliably capturing the full spectral detail available from short TE acquisitions. In addition, the importance of achieving a highly homogenous static magnetic field (B0) in the acquisition region is emphasized, and the limitations of current methods and hardware are discussed. Most recommendations require only software improvements, greatly enhancing the capabilities of clinical MRS on existing hardware. Implementation of these recommendations should strengthen current clinical applications and advance progress toward developing and validating new MRS biomarkers for clinical use.