3D magnetization‐prepared rapid gradient echo with a RF saturation band on a compact 3T scanner
Speltz, Lydia J. Bardwell , In, Myung‐Ho , Trzasko, Joshua D. , Halverson, Maria, Gray, Erin M. , Shu, Yunhong , Bardwell Speltz, Lydia J , Gray, Erin, In, Myung-Ho , Kang, Daehun , Campeau, Norbert G. , Huston, John , Warner, David O. , Bernstein, Matt A.

Medical Physics | August 14, 2025 | Volume: 52 | Issue: 8 | Pages: No page available

10.1002/mp.18039 Get It

Abstract Background Flow‐related ghost artifact from a 3D Magnetization Prepared Rapid Gradient Echo (MPRAGE) sequence results from unsaturated magnetization of incoming arterial flow. This is especially common on MR scanners equipped with smaller radiofrequency (RF) transmit coils. A high‐performance compact 3T (C3T) scanner features a smaller RF‐transmit coil (inner diameter 37 cm, length 40 cm), leading to a rapid fall‐off of the B1 field below the neck. This configuration results in more intense flow‐related ghost artifacts, especially in younger patients. Purpose The C3T scanner's smaller RF‐transmit coil provides RF field coverage over the brain region, causing bright in‐flow arterial signals and flow‐related artifacts in the conventional 3D‐MPRAGE scans. The purpose of this study is to suppress these artifacts by adding a RF saturation band (RFSB) pulses to the 3D‐MPRAGE sequence. Methods The RFSB was added to the 3D‐MPRAGE sequence as a preparation pulse option. To test the effectiveness, 37 subjects were scanned on the C3T under an IRB‐approved protocol using 3D‐MPRAGE with and without RFSB. Ten of those subjects underwent repeated scans with and without RFSB to evaluate test‐retest reliability. A consensus evaluation by two neuroradiologists was performed on all data to compare signal to noise ratio, image contrast, presence of artifacts, and diagnostic confidence. Quantitative analysis included calculating test‐retest differences by image subtraction and evaluating the variance in flow‐artifact‐induced image intensity between the scans with and without RFSB. Additionally, one subject was scanned on a whole‐body 3T scanner using a transmit/receive (T/R) head coil to demonstrate the method's applicability across different MRI platforms as a proof of concept. Results The Wilcoxon signed‐rank test of the neuroradiologist evaluations showed a significant reduction in artifacts and an improvement in diagnostic confidence in the posterior fossa region with and without RFSB ( p < 0.0001). Test–retest analysis showed that adding RFSB significantly reduced image intensity variability in the cerebellum, even among subjects without visible flow artifacts. The normalized difference decreased from 8.71% to 6.41% ( p = 0.0059), suggesting improved image reliability in regions prone to flow‐related artifacts. Additionally, similar findings were observed in scans on a whole‐body 3T with a T/R head coil, demonstrating the broader applicability of this method. Conclusion Incorporating RFSB into 3D‐MPRAGE scans effectively reduces flow‐related ghost artifact on the C3T scanner, improving image quality and diagnostic confidence. These findings suggest that the proposed method could be widely implemented across MRI systems utilizing a smaller RF‐transmit coil.
Modeling and measurement of lead tip heating and resonant length for implanted, insulated wires
Speltz, Lydia J. Bardwell , Shu, Yunhong , Tarasek, Matthew , Trzasko, Joshua D. , Bernstein, Matt A. , Bardwell Speltz, Lydia J , Bernstein, Matt A , Lee, Seung‐Kyun , Foo, Thomas

Magnetic Resonance in Medicine | May 31, 2024 | Volume: 92 | Issue: 4 | Pages: 1714-1727

10.1002/mrm.30145 Get It

AbstractPurposeTo study implant lead tip heating because of the RF power deposition by developing mathematical models and comparing them with measurements acquired at 1.5 T and 3 T, especially to predict resonant length.Theory and MethodsA simple exponential model and an adapted transmission line model for the electric field transfer function were developed. A set of wavenumbers, including that calculated from insulated antenna theory (King wavenumber) and that of the embedding medium were considered. Experiments on insulated, capped wires of varying lengths were performed to determine maximum temperature rise under RF exposure. The results are compared with model predictions from analytical expressions derived under the assumption of a constant electric field, and with those numerically calculated from spatially varying, simulated electric fields from body coil transmission. Simple expressions for the resonant length bounded between one‐quarter and one‐half wavelength are developed based on the roots of transcendental equations.ResultsThe King wavenumber for both models more closely matched the experimental data with a maximum root mean square error of 9.81°C at 1.5 T and 5.71°C at 3 T compared to other wavenumbers with a maximum root mean square error of 27.52°C at 1.5 T and 22.01°C for 3 T. Resonant length was more accurately predicted compared to values solely based on the embedding medium.ConclusionAnalytical expressions were developed for implanted lead heating and resonant lengths under specific assumptions. The value of the wavenumber has a strong effect on the model predictions. Our work could be used to better manage implanted device lead tip heating.