Abstract Objective . Photon-counting detector (PCD) CT enables routine virtual-monoenergetic image (VMI) reconstruction. We evaluated the performance of an automatic VMI energy level (keV) selection tool on a clinical PCD-CT system in comparison to an automatic tube potential (kV) selection tool from an energy-integrating-detector (EID) CT system from the same manufacturer. Approach. Four torso-shaped phantoms (20–50 cm width) containing iodine (2, 5, and 10 mg cc −1 ) and calcium (100 mg cc −1 ) were scanned on PCD-CT and EID-CT. Dose optimization techniques, task-based VMI energy level and tube-potential selection on PCD-CT (CARE keV) and task-based tube potential selection on EID-CT (CARE kV), were enabled. CT numbers, image noise, and dose-normalized contrast-to-noise ratio (CNR d ) were compared. Main results . PCD-CT produced task-specific VMIs at 70, 65, 60, and 55 keV for non-contrast, bone, soft tissue with contrast, and vascular settings, respectively. A 120 kV tube potential was automatically selected on PCD-CT for all scans. In comparison, EID-CT used x-ray tube potentials from 80 to 150 kV based on imaging task and phantom size. PCD-CT achieved consistent dose reduction at 9%, 21% and 39% for bone, soft tissue with contrast, and vascular tasks relative to the non-contrast task, independent of phantom size. On EID-CT, dose reduction factor for contrast tasks relative to the non-contrast task ranged from a 65% decrease (vascular task, 70 kV, 20 cm phantom) to a 21% increase (soft tissue with contrast task, 150 kV, 50 cm phantom) due to size-specific tube potential adaptation. PCD-CT CNR d was equivalent to or higher than those of EID-CT for all tasks and phantom sizes, except for the vascular task with 20 cm phantom, where 70 kV EID-CT CNR d outperformed 55 keV PCD-CT images. Significance . PCD-CT produced more consistent CT numbers compared to EID-CT due to standardized VMI output, which greatly benefits standardization efforts and facilitates radiation dose reduction.
Full field-of-view, high-resolution, photon-counting detector CT: technical assessment and initial patient experience
Abstract We report a comprehensive evaluation of a full field-of-view (FOV) photon-counting detector (PCD) computed tomography (CT) system using phantoms, and qualitatively assess image quality in patient examples. A whole-body PCD-CT system with 50 cm FOV, 5.76 cm z-detector coverage and two acquisition modes (standard: 144 × 0.4 mm collimation and ultra-high resolution (UHR): 120 × 0.2 mm collimation) was used in this study. Phantoms were scanned to assess image uniformity, CT number accuracy, noise power spectrum, spatial resolution, material decomposition and virtual monoenergetic imaging (VMI) performance. Four patients were scanned on the PCD-CT system with matched or lower radiation dose than their prior clinical CT scans performed using energy-integrating detector (EID) CT, and the potential clinical impact of PCD-CT was qualitatively evaluated. Phantom results showed water CT numbers within ±5 HU, and image uniformity measured between peripheral and central regions-of-interests to be within ±5 HU. For the UHR mode using a dedicated sharp kernel, the cut-off spatial frequency was 40 line-pairs cm −1 , which corresponds to a 125 μ m limiting in-plane spatial resolution. The full-width-at-half-maximum for the section sensitivity profile was 0.33 mm for the smallest slice thickness (0.2 mm) using the UHR mode. Material decomposition in a multi-energy CT phantom showed accurate material classification, with a root-mean-squared-error of 0.3 mg cc −1 for iodine concentrations (2–15 mg cc −1 ) and 14.2 mg cc −1 for hydroxyapatite concentrations (200 and 400 mg cc −1 ). The average percent error for CT numbers corresponding to the iodine concentrations in VMI (40–70 keV) was 2.75%. Patient PCD-CT images demonstrated better delineation of anatomy for chest and temporal bone exams performed with the UHR mode, which allowed the use of very sharp kernels not possible with EID-CT. VMI and virtual non-contrast images generated from a patient head CT angiography exam using the standard acquisition mode demonstrated the multi-energy capability of the PCD-CT system.