The Jarvik 2000 bridge to transplant investigational device exemption study was a multicentered, prospective study of 150 UNOS status I patients implanted with the Jarvik 2000 between 2005 and 2012. During the study period, there were numerous modifications of the system that included converting from pin to cone bearings. Results were analyzed for three cohorts: total (n = 150), pin (n = 128), and cone (n = 22). Baseline demographics included age (52 ± 13), gender (79% male), size (BSA 1.98), and etiology (37% idiopathic dilated cardiomyopathy; 43% Ischemic). Seventy percent of patients were either INTERMACS 1 or 2. The primary endpoint—defined as successful transplantation or listing at 180 days (prespecified at 65%; 95% lower CI: 57%)—was successfully achieved for the total cohort (67.3%; 95% CI: 59.5%–74.3%; p = 0.006). In subgroup analysis of the more contemporary, cone-bearing group, the primary endpoint was met in 91% (95% CI: 72%–97.5%; p = 0.001). Compared with pin patients, cone-bearing patients had less hemolysis as well as decreased end-organ dysfunction. Functional and quality of life scores improved after implantation independent type of bearing. In conclusion, despite a particularly sick patient population, the Jarvik 2000 was shown to be effective in supporting the advanced HF patient.
Pump Position Impacts HeartMate II Left Ventricular Assist Device Thrombosis
The PREVENtion of HeartMate II pump Thrombosis through clinical management (PREVENT) study was a multicenter, prospective investigation to evaluate the rate of pump thrombosis (PT) with adoption of a uniform set of surgical and medical practices for left ventricular assist device implantation. We sought to quantify pump position at baseline and retrospectively define a pump position associated with poor clinical outcomes. Chest x-rays at baseline were prospectively obtained per protocol. Pump pocket depth, inflow cannula (IC) angle relative to the pump, and IC angle relative to the vertical were measured. Pumps falling in the tail-ends of the IC angle and pump pocket depth distributions were categorized as having an extreme pump position within the PREVENT study. Patients with extreme pump position had a significantly higher risk of confirmed and suspected PT, hemolysis, and elevated lactate dehydrogenase. In a multivariable analysis of survival free of confirmed PT, extreme pump position was an independent risk factor (hazard ratio = 3.6; 95% confidence interval = 1.5–8.9; p = 0.006) when adjusting for differences in pump speed and anticoagulation level. Our analysis shows that HeartMate II pump position at implant can significantly impact event-free survival and the incidence of adverse events at 6 months.