Back to Authors

Korinek, Sarah

Mayo Clinic

Mayo Clinic

Citations 1

Kinematics of semi-constrained total elbow arthroplasty
An, K-N., Korinek, Sarah, Morrey, BF, O’Driscoll, Shawn W. , Morrey, BF

The Journal of Bone and Joint Surgery. British volume | February 21, 2018 | Volume: 74-B | Issue: 2 | Pages: 297-299

10.1302/0301-620x.74b2.1544973 Get It

We used 11 cadaver elbows and a three-dimensional electromagnetic tracking device to record elbow movements before and after implantation of a 'loose-hinged' elbow prosthesis (modified Coonrad). During simulated active motion there was a maximum of 2.7 degrees (+/- 1.5 degrees) varus/valgus laxity in the cadaver joints. This increased slightly after total elbow arthroplasty to 3.8 degrees (+/- 1.4 degrees). These values are lower than those recorded for the cadaver joints and for the prostheses at the limits of their varus/valgus displacements, indicating that both behave as 'semi-constrained' joints under physiological conditions. They suggest that the muscles absorb some of the forces and moments that in a constrained prosthesis would be transferred to the prosthesis-bone interface.