BACKGROUND: Quantitative susceptibility mapping (QSM) and dynamic contrast-enhanced quantitative perfusion (DCEQP) magnetic resonance imaging sequences assessing iron deposition and vascular permeability were previously correlated with new hemorrhage in cerebral cavernous malformations. We assessed their prospective changes in a multisite trial-readiness project. METHODS: Patients with cavernous malformation and symptomatic hemorrhage (SH) in the prior year, without prior or planned lesion resection or irradiation were enrolled. Mean QSM and DCEQP of the SH lesion were acquired at baseline and at 1- and 2-year follow-ups. Sensitivity and specificity of biomarker changes were analyzed in relation to predefined criteria for recurrent SH or asymptomatic change. Sample size calculations for hypothesized therapeutic effects were conducted. RESULTS: We logged 143 QSM and 130 DCEQP paired annual assessments. Annual QSM change was greater in cases with SH than in cases without SH ( P =0.019). Annual QSM increase by ≥6% occurred in 7 of 7 cases (100%) with recurrent SH and in 7 of 10 cases (70%) with asymptomatic change during the same epoch and 3.82× more frequently than clinical events. DCEQP change had lower sensitivity for SH and asymptomatic change than QSM change and greater variance. A trial with the smallest sample size would detect a 30% difference in QSM annual change during 2 years of follow-up in 34 or 42 subjects (1 and 2 tailed, respectively); power, 0.8, α=0.05. CONCLUSIONS: Assessment of QSM change is feasible and sensitive to recurrent bleeding in cavernous malformations. Evaluation of an intervention on QSM percent change may be used as a time-averaged difference between 2 arms using a repeated measures analysis. DCEQP change is associated with lesser sensitivity and higher variability than QSM. These results are the basis of an application for certification by the US Food and Drug Administration of QSM as a biomarker of drug effect on bleeding in cavernous malformations. REGISTRATION: URL: https://www.clinicaltrials.gov ; Unique identifier: NCT03652181.
Phantom validation of quantitative susceptibility and dynamic contrast‐enhanced permeability MR sequences across instruments and sites
BackgroundQuantitative susceptibility mapping (QSM) and dynamic contrast‐enhanced quantitative permeability (DCEQP) on magnetic resonance (MR) have been shown to correlate with neurovascular disease progression as markers of vascular leakage and hemosiderin deposition. Applying these techniques as monitoring biomarkers in clinical trials will be necessary; however, their validation across multiple MR platforms and institutions has not been rigorously verified.PurposeTo validate quantitative measurement of MR biomarkers on multiple instruments at different institutions.Study TypePhantom validation between platforms and institutions.Phantom ModelT1/susceptibility phantom, two‐compartment dynamic flow phantom.Field Strength/Sequence3T/QSM, T1 mapping, dynamic 2D SPGR.AssessmentPhilips Ingenia, Siemens Prisma, and Siemens Skyra at three different institutions were assessed. A QSM phantom with concentrations of gadolinium, corresponding to magnetic susceptibilities of 0, 0.1, 0.2, 0.4, and 0.8 ppm was assayed. DCEQP was assessed by measuring a MultiHance bolus as the consistency of the width ratio of the curves at the input and outputs over a range of flow ratios between outputs.Statistical TestsEach biomarker was assessed by measures of accuracy (Pearson correlation), precision (paired t‐test between repeated measurements), and reproducibility (analysis of covariance [ANCOVA] between instruments).ResultsQSM accuracy of r2 > 0.997 on all three platforms was measured. Precision (P = 0.66 Achieva, P = 0.76 Prisma, P = 0.69 Skyra) and reproducibility (P = 0.89) were good. T1 mapping of accuracy was r2 > 0.98. No significant difference between width ratio regression slopes at site 2 (P = 0.669) or site 3 (P = 0.305), and no significant difference between width ratio regression slopes between sites was detected by ANCOVA (P = 0.48).Data ConclusionThe phantom performed as expected and determined that MR measures of QSM and DCEQP are accurate and consistent across repeated measurements and between platforms.Level of Evidence: 1Technical Efficacy Stage: 2J. Magn. Reson. Imaging 2020;51:1192–1199.