Time structures of proton pencil beam scanning delivery on a microsecond scale measured with a pixelated semiconductor detector Timepix3
Ding, Xuanfeng , Liang, Xiaoying , Oancea, Cristina , Wang, Peilong, Rule, William G., Liu, Wei , Shen, Jiajian , Liang, Xiaoying , Liu, Wei , Shen, Jiajian , Charyyev, Serdar , Wang, Peilong, Bues, Martin , Lin, Liyong

Journal of Applied Clinical Medical Physics | August 13, 2024 | Volume: 25 | Issue: 9 | Pages: No page available

10.1002/acm2.14486 Get It

AbstractPurposeThe time structures of proton spot delivery in proton pencil beam scanning (PBS) radiation therapy are essential in many clinical applications. This study aims to characterize the time structures of proton PBS delivered by both synchrotron and synchrocyclotron accelerators using a non‐invasive technique based on scattered particle tracking.MethodsA pixelated semiconductor detector, AdvaPIX‐Timepix3, with a temporal resolution of 1.56 ns, was employed to measure time of arrival of secondary particles generated by a proton beam. The detector was placed laterally to the high‐flux area of the beam in order to allow for single particle detection and not interfere with the treatment. The detector recorded counts of radiation events, their deposited energy and the timestamp associated with the single events. Individual recorded events and their temporal characteristics were used to analyze beam time structures, including energy layer switch time, magnet switch time, spot switch time, and the scanning speeds in the x and y directions. All the measurements were repeated 30 times on three dates, reducing statistical uncertainty.ResultsThe uncertainty of the measured energy layer switch times, magnet switch time, and the spot switch time were all within 1% of average values. The scanning speeds uncertainties were within 1.5% and are more precise than previously reported results. The measurements also revealed continuous sub‐milliseconds proton spills at a low dose rate for the synchrotron accelerator and radiofrequency pulses at 7 µs and 1 ms repetition time for the synchrocyclotron accelerator.ConclusionThe AdvaPIX‐Timepix3 detector can be used to directly measure and monitor time structures on microseconds scale of the PBS proton beam delivery. This method yielded results with high precision and is completely independent of the machine log files.
Impact of proton PBS machine operating parameters on the effectiveness of layer rescanning for interplay effect mitigation in lung SBRT treatment
Liang, Xiaoying , Flampouri, Stella , Park, Justin C., Yaddanapudi, Sridhar , Tan, Jun , Beltran, Chris , Tan, Jun , Shen, Jiajian , Yaddanapudi, Sridhar , Beltran, Chris J , Liang, Xiaoying , Liu, Chunbo , Shen, Jiajian , Lu, Bo, Furutani, Keith M. , Tan, Jun

Journal of Applied Clinical Medical Physics | April 9, 2024 | Volume: 25 | Issue: 7 | Pages: No page available

10.1002/acm2.14342 Get It

Abstract Background Rescanning is a common technique used in proton pencil beam scanning to mitigate the interplay effect. Advances in machine operating parameters across different generations of particle therapy systems have led to improvements in beam delivery time (BDT). However, the potential impact of these improvements on the effectiveness of rescanning remains an underexplored area in the existing research. Methods We systematically investigated the impact of proton machine operating parameters on the effectiveness of layer rescanning in mitigating interplay effect during lung SBRT treatment, using the CIRS phantom. Focused on the Hitachi synchrotron particle therapy system, we explored machine operating parameters from our institution's current (2015) and upcoming systems (2025A and 2025B). Accumulated dynamic 4D dose were reconstructed to assess the interplay effect and layer rescanning effectiveness. Results Achieving target coverage and dose homogeneity within 2% deviation required 6, 6, and 20 times layer rescanning for the 2015, 2025A, and 2025B machine parameters, respectively. Beyond this point, further increasing the number of layer rescanning did not further improve the dose distribution. BDTs without rescanning were 50.4, 24.4, and 11.4 s for 2015, 2025A, and 2025B, respectively. However, after incorporating proper number of layer rescanning (six for 2015 and 2025A, 20 for 2025B), BDTs increased to 67.0, 39.6, and 42.3 s for 2015, 2025A, and 2025B machine parameters. Our data also demonstrated the potential problem of false negative and false positive if the randomness of the respiratory phase at which the beam is initiated is not considered in the evaluation of interplay effect. Conclusion The effectiveness of layer rescanning for mitigating interplay effect is affected by machine operating parameters. Therefore, past clinical experiences may not be applicable to modern machines.
Investigation of scan path optimization in improving proton pencil beam scanning continuous delivery
Liu, Chunbo , Beltran, Chris , Park, Chunjoo, Yaddanapudi, Sridhar , Tan, Jun , Furutani, Keith M. , Liang, Xiaoying , Tan, Jun , Shen, Jiajian , Furutani, Keith M , Liang, Xiaoying , Beltran, Chris J , Yaddanapudi, Sridhar , Shen, Jiajian , Lu, Bo, Tan, Jun

Physics in Medicine & Biology | September 13, 2023 | Volume: 68 | Issue: 19 | Pages: 195023

10.1088/1361-6560/acf98e Get It

Abstract Objective . To investigate the impact of scan path optimization on the dose accuracy and beam delivery time (BDT) of proton pencil beam scanning in the dose-driven continuous scanning (DDCS). Approach . A diverse set of six clinical plans, representing various spot patterns and treatment sites, was used to evaluate the effectiveness of scan time optimization and scan length optimization. The DDCS dose discrepancy and BDT with optimized scan paths was compared to the default serpentine scan path. Main results . Both scan time optimization and scan path optimization were able to reduce the DDCS dose discrepancy compared to the default serpentine scan path. All plans, except for the layer repainting lung plan, achieved a 2%/2 mm gamma pass rate of over 99% and less than 1% PTV DVH root mean square error (RMSE) through scan path optimization. In the case of the layer repainting lung plan, when compared to the default serpentine scan path, the 2%/2 mm gamma pass rate showed improvements from 91.3% to 93.1% and 95.8%, while the PTV DVH RMSE decreased from 2.1% to 1.7% and 1.1% for scan time optimization and scan length optimization, respectively. Although scan time optimization resulted in shorter total scan times for all plans compared to the default scan path and scan length optimization tended to have longer total scan times. However, due to the short total scan times and their minimal contribution to the total BDT, the impact of scan path optimization on the total BDT was practically negligible. Significance . Both scan time optimization and scan length optimization proved to be effective in minimizing DDCS dose discrepancy. No definitive winner can be determined between these two optimization approaches. Both scan time and scan length optimization had minimal effect on the total BDT.
A Geant4 shielding design for the first US carbon multi-ion hybrid synchrotron facility
Dougherty, Jingjing M. , Furutani, Keith M. , Guatelli, Susanna , Liang, Xiaoying , Rosenfeld, Anatoly , Beltran, Chris , Furutani, Keith M , Guatelli, S , Liang, Xiaoying , Rosenfeld, A B , Beltran, Chris J , Dougherty, Jingjing M , Bolst, David

Physics in Medicine & Biology | February 2, 2023 | Volume: 68 | Issue: 5 | Pages: 054002

10.1088/1361-6560/acb887 Get It

Abstract The Mayo Clinic Florida Integrated Oncology Building will be the home of the first spot-scanning only carbon/proton hybrid therapy system by Hitachi, Ltd. It will provide proton beams up to kinetic energies of 230 MeV and carbon beams up to 430 MeV n −1 for clinical deployment. To provide adequate radiation protection, the Geant4 (v10.6) Monte Carlo toolkit was utilized to quantify the ambient dose equivalent at a 10 mm depth ( H *(10)) for photons and neutrons. To perform accurate calculations of the ambient dose equivalent, three-dimensional computer-aided design files of the entire planned facility were imported into Geant4, as well as certain particle system components such as the bending magnets, fast Faraday cup, and gantry. Particle fluence was scored using 60 cm diameter spheres, which were strategically placed throughout areas of interests. Analytical calculations were performed as first-pass design checks. Major shielding slabs were optimized using Geant4 simulations iteratively, with more than 20 alternative designs evaluated within Geant4. The 430 MeV n −1 carbon beams played the most significant role in concrete thickness Requirements. The primary wall thickness for the carbon fixed beam room is 4 meters. The presence of the proton gantry structure in the simulation caused the ambient dose equivalent to increase by around 67% at the maze entrance, but a decrease in the high energy beam transport corridor. All shielding primary and secondary goals for clinical operations were met per state regulation and national guidelines.
Investigation of beam delivery time for synchrotron-based proton pencil beam scanning system with novel scanning mode
Liang, Xiaoying , Liu, Chunbo , Shen, Jiajian , Bues, Martin , Dougherty, Jingjing M. , Parisi, Alessio , Shrestha, Deepak K. , Yaddanpudi, Sridhar, Beltran, Chris , Yaddanapudi, Sridhar , Furutani, Keith M. , Shen, Jiajian , Bues, Martin , Dougherty, Jingjing M , Parisi, Alessio , Beltran, Chris J , Liang, Xiaoying , Yaddanapudi, Sridhar , Li, Heng

Physics in Medicine & Biology | July 25, 2022 | Volume: 67 | Issue: 17 | Pages: 175001

10.1088/1361-6560/ac8410 Get It

Abstract Objective. To investigate synchrotron-based proton pencil beam scanning (PBS) beam delivery time (BDT) using novel continuous scanning mode. Approach. A BDT calculation model was developed for the Hitachi particle therapy system. The model was validated against the measured BDT of 36 representative clinical proton PBS plans with discrete spot scanning (DSS) in the current Hitachi proton therapy system. BDTs were calculated with the next generation using Mayo Clinic Florida system operating parameters for conventional DSS, and novel dose driven continuous scanning (DDCS). BDTs of DDCS with and without Break Spots were investigated. Main results. For DDCS without Break Spots, the use of Stop Ratio to control the transit dose largely reduced the beam intensity and consequently, severely prolonged the BDT. DDCS with Break Spots was able to maintain a sufficiently high beam intensity while controlling transit dose. In DDCS with Break Spots, tradeoffs were made between beam intensity and number of Break Spots. Therefore, BDT decreased with increased beam intensity but reached a plateau for beam intensity larger than 10 MU s −1 . Averaging over all clinical plans, BDT was reduced by 10% for DDCS with Break Spots compared to DSS. Significance. DDCS with Break Spots reduced BDT. DDCS has the potential to further reduce BDT under the ideal scenario which requests both stable beam intensity extraction and accurately modelling the transit dose. Further investigation is warranted.