Pulsed Field Ablation for the Treatment of Atrial Fibrillation: PULSED AF Pivotal Trial
Haines, David E. , Boersma, Lucas V.A. , Sood, Nitesh , Natale, Andrea , Marchlinski, Francis E. , Calkins, Hugh , Sanders, Prashanthan , Packer, Douglas L. , Kuck, Karl-Heinz , Hindricks, Gerhard , Önal, Birce , Cerkvenik, Jeffrey , Tada, Hiroshi , Investigators, on behalf of the PULSED AF, Verma, Atul , Boersma, Lucas , Natale, Andrea , Marchlinski, Francis E , Calkins, Hugh , Sanders, Prashanthan , Packer, Douglas L , Kuck, Karl-Heinz , Onal, Birce , Verma, Atul , DeLurgio, David B. , Investigators, on behalf of the PULSED AF

Circulation | March 6, 2023 | Volume: 147 | Issue: 19 | Pages: 1422-1432

10.1161/circulationaha.123.063988 Get It

Background: Pulsed field ablation uses electrical pulses to cause nonthermal irreversible electroporation and induce cardiac cell death. Pulsed field ablation may have effectiveness comparable to traditional catheter ablation while preventing thermally mediated complications. Methods: The PULSED AF pivotal study (Pulsed Field Ablation to Irreversibly Electroporate Tissue and Treat AF) was a prospective, global, multicenter, nonrandomized, paired single-arm study in which patients with paroxysmal (n=150) or persistent (n=150) symptomatic atrial fibrillation (AF) refractory to class I or III antiarrhythmic drugs were treated with pulsed field ablation. All patients were monitored for 1 year using weekly and symptomatic transtelephonic monitoring; 3-, 6-, and 12-month ECGs; and 6- and 12-month 24-hour Holter monitoring. The primary effectiveness end point was freedom from a composite of acute procedural failure, arrhythmia recurrence, or antiarrhythmic escalation through 12 months, excluding a 3-month blanking period to allow recovery from the procedure. The primary safety end point was freedom from a composite of serious procedure- and device-related adverse events. Kaplan-Meier methods were used to evaluate the primary end points. Results: Pulsed field ablation was shown to be effective at 1 year in 66.2% (95% CI, 57.9 to 73.2) of patients with paroxysmal AF and 55.1% (95% CI, 46.7 to 62.7) of patients with persistent AF. The primary safety end point occurred in 1 patient (0.7%; 95% CI, 0.1 to 4.6) in both the paroxysmal and persistent AF cohorts. Conclusions: PULSED AF demonstrated a low rate of primary safety adverse events (0.7%) and provided effectiveness consistent with established ablation technologies using a novel irreversible electroporation energy to treat patients with AF. Registration: URL: https://www.clinicaltrials.gov ; Unique identifier: NCT04198701.
First-in-Human Experience and Acute Procedural Outcomes Using a Novel Pulsed Field Ablation System: The PULSED AF Pilot Trial
Boersma, Lucas V.A. , Haines, David E., Natale, Andrea , Marchlinski, Francis E. , Sanders, Prashanthan , Calkins, Hugh , Packer, Douglas L. , Hummel, John D. , Önal, Birce , Rosen, Sofi, Kuck, Karl-Heinz , Hindricks, Gerhard, Wilsmore, Bradley , Verma, Atul , Boersma, Lucas , Natale, Andrea , Marchlinski, Francis E , Sanders, Prashanthan , Calkins, Hugh , Packer, Douglas L , Hummel, John , Onal, Birce , Kuck, Karl-Heinz , Wilsmore, Bradley , Haines, David E., Rosen, Sofi, Hindricks, Gerhard

Circulation: Arrhythmia and Electrophysiology | December 29, 2021 | Volume: 15 | Issue: 1 | Pages: No page available

10.1161/circep.121.010168 Get It

Background: Pulsed field ablation (PFA) is a novel form of ablation using electrical fields to ablate cardiac tissue. There are only limited data assessing the feasibility and safety of this type of ablation in humans. Methods: PULSED AF (Pulsed Field Ablation to Irreversibly Electroporate Tissue and Treat AF; https://www.clinicaltrials.gov ; unique identifier: NCT04198701) is a nonrandomized, prospective, multicenter, global, premarket clinical study. The first-in-human pilot phase evaluated the feasibility and efficacy of pulmonary vein isolation using a novel PFA system delivering bipolar, biphasic electrical fields through a circular multielectrode array catheter (PulseSelect; Medtronic, Inc). Thirty-eight patients with paroxysmal or persistent atrial fibrillation were treated in 6 centers in Australia, Canada, the United States, and the Netherlands. The primary outcomes were ability to achieve acute pulmonary vein isolation intraprocedurally and safety at 30 days. Results: Acute electrical isolation was achieved in 100% of pulmonary veins (n=152) in the 38 patients. Skin-to-skin procedure time was 160±91 minutes, left atrial dwell time was 82±35 minutes, and fluoroscopy time was 28±9 minutes. No serious adverse events related to the PFA system occurred in the 30-day follow-up including phrenic nerve injury, esophageal injury, stroke, or death. Conclusions: In this first-in-human clinical study, 100% pulmonary vein isolation was achieved using only PFA with no PFA system–related serious adverse events. Graphic Abstract: A graphic abstract is available for this article.
Reduction in Pulmonary Vein Stenosis and Collateral Damage With Pulsed Field Ablation Compared With Radiofrequency Ablation in a Canine Model
Howard, Brian, Haines, David E. , Packer, Douglas L. , Kirchhof, Nicole , Barka, Noah , Önal, Birce , Fraasch, Steve, Miklavčič, Damijan , Stewart, Mark T. , Packer, Douglas, Onal, Birce , Verma, Atul

Circulation: Arrhythmia and Electrophysiology | July 27, 2020 | Volume: 13 | Issue: 9 | Pages: No page available

10.1161/circep.120.008337 Get It

Background: Pulmonary vein (PV) stenosis is a highly morbid condition that can result after catheter ablation for PV isolation. We hypothesized that pulsed field ablation (PFA) would reduce PV stenosis risk and collateral injury compared with irrigated radiofrequency ablation (IRF). Methods: IRF and PFA deliveries were randomized in 8 dogs with 2 superior PVs ablated using one technology and 2 inferior PVs ablated using the other technology. IRF energy (25–30 W) or PFA was delivered (16 pulse trains) at each PV in a proximal and in a distal site. Contrast computed tomography scans were collected at 0, 2, 4, 8, and 12-week (termination) time points to monitor PV cross-sectional area at each PV ablation site. Results: Maximum average change in normalized cross-sectional area at 4-weeks was −46.1±45.1% post-IRF compared with −5.5±20.5% for PFA ( P ≤0.001). PFA-treated targets showed significantly fewer vessel restrictions compared with IRF ( P ≤0.023). Necropsy showed expansive PFA lesions without stenosis in the proximal PV sites, compared with more confined and often incomplete lesions after IRF. At the distal PV sites, only IRF ablations were grossly identified based on focal fibrosis. Mild chronic parenchymal hemorrhage was noted in 3 left superior PV lobes after IRF. Damage to vagus nerves as well as evidence of esophagus dilation occurred at sites associated with IRF. In contrast, no lung, vagal nerve, or esophageal injury was observed at PFA sites. Conclusions: PFA significantly reduced risk of PV stenosis compared with IRF postprocedure in a canine model. IRF also caused vagus nerve, esophageal, and lung injury while PFA did not.