Back to Authors

Engle, Dean D.

Mayo Clinic

Mayo Clinic

Citations 2

Oversensing of transthoracic excitation stimuli in contemporary pacemakers
Ryan, J. , Tempel, Nathan D., Engle, Dean D., Hayes, David L. , Cha, Yong‐Mei , Asirvatham, Samuel J. , Ryan, James D , Asirvatham, Samuel J

Pacing and Clinical Electrophysiology | February 1, 2018 | Volume: 41 | Issue: 3 | Pages: 340-340

10.1111/pace.13290 Get It

Oversensing of transthoracic excitation stimuli in contemporary pacemakers
Ryan, J. , Tempel, Nathan D., Engle, Dean D., Hayes, David L. , Cha, Yong‐Mei , Asirvatham, Samuel J. , Ryan, James D , Asirvatham, Samuel J

Pacing and Clinical Electrophysiology | January 10, 2018 | Volume: 41 | Issue: 2 | Pages: 161-166

10.1111/pace.13269 Get It

AbstractBackgroundTransthoracic impedance measurements (TIM) is primarily used for minute ventilation rate adaptive sensors in pacemakers. With elevated impedance, the TIM electrical signal itself is oversensed, causing device malfunction.ObjectiveWe report an increased incidence of TIM oversensing.MethodRetrospective chart review. We review existing records of 18 patients who have demonstrated device malfunction with TIM oversensing.ResultsWe have found a 1.8% incidence of TIM‐related oversensing in our patient population of 959 patients with contemporary Boston Scientific (Marlborough, MA, USA) pacemakers and cardiac resynchronization therapy pacemakers. One patient experienced a syncopal episode.ConclusionOversensing with pacing inhibition is apparent with the potential of adverse effects to patients.