Sequencing and Resolution of Amplified Herpes Simplex Virus DNA with Intermediate Melting Curves as Genotype 1 or 2 by LightCycler PCR Assay
Issa, Nicolas C. , Espy, Mark J., Uhl, James R., Smith, Thomas F.

Journal of Clinical Microbiology | April 6, 2005 | Volume: 43 | Issue: 4 | Pages: 1843-1845

10.1128/jcm.43.4.1843-1845.2005 Get It

ABSTRACT DNA from 101 specimens containing herpes simplex virus (HSV) produced atypical intermediate melting curves compared with those expected for HSV type 1 or HSV type 2 subsequent to real-time PCR. Nucleic acid sequence analysis of amplified target DNA revealed 1- or 3-bp polymorphisms in the probe region which allowed designation of these viruses as HSV genotype 1 or HSV genotype 2. These two subpopulations of HSV were also identified as HSV genotype 1 or HSV genotype 2 using another commercially available PCR method. Amplified HSV target DNA producing intermediate melting curves could be designated as HSV genotype 1 or HSV genotype 2 without performing sequencing or another PCR method with 96/101 (95%) specimens by adding known intermediate HSV DNA characteristic for the two subpopulations as controls.
Comparison of Specimen Processing and Nucleic Acid Extraction by the Swab Extraction Tube System versus the MagNA Pure LC System for Laboratory Diagnosis of Herpes Simplex Virus Infections by LightCycler PCR
Issa, Nicolas C. , Espy, Mark J., Uhl, J. R., Harmsen, William S. , Mandrekar, Jayawant N. , Gullerud, R. E., Davis, Mike , Smith, Thomas F.

Journal of Clinical Microbiology | March 4, 2005 | Volume: 43 | Issue: 3 | Pages: 1059-1063

10.1128/jcm.43.3.1059-1063.2005 Get It

ABSTRACT A total of 563 specimens (234 dermal and 329 genital swabs) from patients suspected of having herpes simplex virus (HSV) infections were processed using two different extraction methods (the MagNA Pure LC system and the swab extraction tube system [SETS]); HSV DNA was amplified by LightCycler PCR. HSV DNA was detected in 157 of 563 specimens (27.9%) processed by the MagNA Pure LC system and in 179 of 563 specimens (31.8%) processed by SETS ( P < 0.0001). There was no specimen processed by the MagNA Pure LC extraction method that was positive only for HSV DNA. Of 157 specimens positive by both methods, HSV DNA copy levels were higher (using cycle crossover points [cycle threshold { C T }]) with SETS (mean C T , 25.9 cycles) than with the MagNA Pure LC system (mean C T , 32.0 cycles) ( P < 0.0001). The time to process 32 samples was longer with the MagNA Pure LC extraction system (90 min) than with SETS (35 min). HSV DNA extraction using SETS is faster, less expensive, and more sensitive than the MagNA Pure LC system and could replace the latter for the laboratory diagnosis of HSV infections using LightCycler PCR.