Five-year outcomes of transcatheter mitral valve replacement in patients with severe symptomatic mitral regurgitation: results from the Tendyne Expanded Clinical Study
Abstract Aims Transcatheter mitral valve replacement (TMVR) with the Tendyne Mitral Valve System is a treatment option for patients with severe symptomatic mitral regurgitation (MR) unsuitable for conventional mitral valve surgery or transcatheter edge-to-edge repair (TEER). This study sought to evaluate the safety and effectiveness of TMVR through 5-year follow-up. Methods and Results The Tendyne Expanded Clinical Study is a prospective, single-arm, multicentre study that enrolled patients between November 2014 and June 2020. The study enrolled 191 patients (mean age 74.1 ± 8.0 years, 62.8% male, 70.2% New York Heart Association Class (NYHA) class III/IV, 88.5% secondary MR), of whom 186 (97.4%) underwent TMVR with Tendyne. MR grade decreased from ≥3+ in 99.5% patients at baseline to grade ≤1+ in 95.3% at 5 years. In those patients surviving to 5 years (n = 49), durable symptomatic improvement was evidenced by 73.5% patients being in NYHA class I/II at 5 years. The Kansas City Cardiomyopathy Questionnaire overall summary score increased from 48.5 ± 22.5 points at baseline to 67.6 ± 21.5 points at 5 years. Serious adverse events that occurred through 5 years included life-threatening bleeding (11.5%), fatal bleeding (2.6%), renal insufficiency/failure (28.8%), endocarditis (7.3%), device thrombosis (5.8%), and new onset atrial fibrillation (13.1%). No structural device degeneration or device embolism occurred through 5 years. Conclusions The Tendyne TMVR was effective at achieving immediate and sustainable elimination of MR, which was associated with symptomatic improvement and absence of structural valve degeneration in a high-risk cohort over 5 years. These findings support TMVR with the Tendyne System as an alternative for patients with appropriate mitral valve anatomy and symptomatic secondary MR unsuitable for mitral surgery or TEER.