BACKGROUND AND OBJECTIVES: Embolization is part of multidisciplinary brain arteriovenous malformations (AVMs) treatment, yet safety data for novel liquid embolization agents (LEAs) such as Squid and PHIL remain limited. This study reports complications and outcomes from curative embolization using Onyx, Squid, PHIL, and Glue. METHODS: Adult patients with brain AVMs treated with LEAs from July 2015 to December 2022 were reviewed for intraprocedural technical issues, postoperative hemorrhagic, and ischemic complications. Minor (modified Rankin Scale ≤2) and major complication rate (morbidity/modified Rankin Scale >2), mortality, and 6-month angiographic cure rate were evaluated. RESULTS: In total, 135 patients (62.2% male, mean age 41.7 ± 15.1 years) with 206 embolizations (34% Onyx, 18.9% Squid, 17.5% PHIL, 28.2% Glue) were included. Overall, 87.4% procedures were in ruptured AVMs and 84% treated by the transarterial technique (84%). The 6-month angiographic cure rate was 53.3%, higher when nonadhesive LEAs were used than Glue (65.3% vs 17.6%, P < .001). In adjusted analysis, only eloquent location (odds ratio [OR] = 4.4; P = .007) and Glue (OR = 9.1; 95% CI 2.9-28.9, P < .001) were predictors of lower angiographic occlusion rate, whereas size had an inverse correlation (OR = 0.67; P = .019). Technical procedural complications, hemorrhagic, and ischemic rates were 10.2% (2.2% morbidity), 13.1% (3.7% morbidity and 2.2% mortality), and 9.2% (2.2% morbidity, 0% mortality), respectively, regardless of the preferred LEA. Minor, major complications, and mortality rates were 3.7%, 4.4%, and 2.2%, respectively. Hemorrhage was the major contributor to morbidity and mortality, whereas deep venous drainage was an independent predictor of hemorrhage and mortality. CONCLUSION: PHIL and Squid demonstrated complication and angiographic cure rates comparable with Onyx. However, curative embolization of brain AVMs resulted in only moderate angiographic occlusion. Larger multicenter studies with standardized operator protocols, cost-effectiveness assessments, and extended follow-up are warranted to validate these findings.
Flow diverter braid deformation following treatment of cerebral aneurysms: incidence, clinical relevance, and potential risk factors
Background Although flow diverters (FDs) have benefited from several technical improvements, recently concerns have arisen regarding the braid stability after implantation. Thus, we investigated frequency, predictive factors, and clinical impact of the phenomenon of FD braid deformation (FDBD). Methods Consecutive intracranial aneurysms (IAs) treated with various FDs, between January 2018 and July 2023, were reviewed to identify FDBD (defined as the deformation of a FD without any external force applied to it). Patient, aneurysm, procedural, and FD characteristics were retrieved and analyzed using univariate and multivariable analyses. Morbidity is defined as a score of +1 in the modified Rankin Scale at 3 months. Results In total, 245 FD procedures (271 FDs implanted; 25 multiple IAs treated with 1 FD) in 228 patients; FDBD was observed in 36/245 cases (14.7%), mainly at follow-up angiography (32/36, 88.9%); fish-mouthing was the most frequent FDBD. Morbidity was related to fish-mouthing and braid collapse and was significantly higher in the FDBD group after retreatment (p=0.04). Drawn filled tubing with platinum (DFT) (adjusted odds ratio (aOR)=7.0, 95% CI 3.0 to 17.5; p<0.001) and FD diameter (aOR=2.2, 95% CI 1.3 to 4.1; p<0.01) were identified as independent predictors of FDBD. The metal alloy composing the FD (p=0.13) and coated surfaces were not significantly associated with FDBD (p=0.54 in multivariable analysis). Conclusions FDBD is a frequent phenomenon observed in about 15% of cases, and it was responsible for higher morbidity. Only FD characteristics (DFT and FD diameter) were independent determinants of FDBD. Future research should focus on the impact of novel braid configurations and materials on braid stability.
Neck apposition is a key factor for aneurysm occlusion after Woven EndoBridge device embolization
Background Sealing of the aneurysm neck with a Woven EndoBridge (WEB) device is recommended for disrupting the blood flow inside the aneurysm. This study investigates the relationship between WEB neck apposition and aneurysm occlusion rates. Methods Aneurysms treated with a WEB from March 2017 to May 2022 at a single center were included. WEB neck apposition (poor/good) and WEB protrusion (yes/no) were evaluated on post-detachment high resolution cone beam CT images. Angiographic occlusion was assessed with the Bicêtre Occlusion Scale score (BOSS). Univariate and multivariable analysis tested the association between neck apposition and occlusion rates. Results The study included 159 aneurysms in 141 patients (mean age 55.8±11.2 years; 64.2% women). Good neck apposition and protrusion were noted in 123 (77.4%) and 30 (18.9%) cases, respectively. Inter-rater agreements were good for neck apposition (κ=0.75) and protrusion (κ=0.78). Complete and adequate occlusion was achieved in 104 (65%) and 130 (82%) cases, respectively (median follow-up 18 months). Good neck apposition was a strong independent predictor for both adequate (adjusted OR (aOR)=5.9, 95% CI 2.4 to 14.9; P<0.001) and complete occlusion (aOR=7.1, 95% CI 3.0 to 18.1; P<0.001). Protrusion was more frequent in the adequate occlusion group versus the aneurysm recurrence group without reaching statistical significance (P=0.06), but was associated with more thromboembolic complications (9/30 (30%) vs 12/129 (9%); P<0.01). WEB shape modification was significantly greater in poor apposition cases (P=0.03). Conclusions Achieving good neck apposition of the WEB strongly predicts aneurysm occlusion during follow-up. WEB protrusion should be minimized due to the increase in thromboembolic risk with limited impact on aneurysm occlusion.