Kelley, Brandon

Mayo Clinic

Mayo Clinic

Citations 5

B-030 Establishment of Concentration-Specific Hemolysis Thresholds for the New Roche Bilirubin Direct Gen.2 Assay and Investigation of Dilution as a Possible Mitigation Strategy for Hemolyzed Specimens
Kelley, Brandon, Lasho, Michael A., Wockenfus, Amy M., Baumann, N, Block, Darci R , Katzman, Brooke M.

Clinical Chemistry | October 2, 2025 | Volume: 71 | Issue: Supplement_1 | Pages: No page available

10.1093/clinchem/hvaf086.428 Get It

Abstract Background Hemolysis, a common pre-analytical interferent in clinical laboratory testing, can significantly affect accuracy and reliability of laboratory results. One test particularly susceptible to interference is direct bilirubin (D-bil). For example, during December 2024, 40% (500/1239) of specimens rejected due to hemolysis were orders for D-bil. The recently released Roche (Roche Diagnostics) direct bilirubin reagent (BILD2) has a lower threshold for hemolysis (H-index [HI]=25)) than the previous generation (HI=30). The aims of our study were to (1) establish concentration-specific hemolysis thresholds for D-bil and (2) investigate the instrument’s reduced volume function (dilution) as a possible mitigation strategy for samples with hemolysis indices exceeding the established thresholds. Methods Residual serum specimens from physician-ordered D-bil testing were used and analyzed on Roche cobas c701 and c501 analyzers. The instrument’s decreased volume function, intended for results outside the measuring range, was used as a 1:2 dilution and validated using 9 non-hemolyzed negative control specimens with D-bil results from 9.8-13.4 mg/dL. Recovery within 10% neat value was considered acceptable. EDTA whole blood was used to prepare hemolysate pools by washing red cells and lysing by freezing (-80 °C). Twenty-six serum pools with varying D-bil concentrations (0.2-3.4 mg/dL) were prepared. Each pool was split into 10 aliquots and spiked with increasing amounts of hemolysate, not exceeding 10% serum to interferent ratio. The decreased volume function (dilution) was utilized on a subset of the hemolysate-spiked aliquots (N=55) with D-bil results from 0.2-2.5 mg/dL. HI values were estimated to be 50% of neat HI due to the instrument’s inability to measure indices on specimens tested with the decreased volume function. Acceptance criteria for experiments utilizing hemolyzed specimens were ±0.1 mg/dL or 20% of neat sample. Results All non-hemolyzed specimens tested with the decreased volume function passed within acceptance criteria with average recovery of 109% (range: 106-110%) from undiluted specimens. Serum pools with D-bil <0.3 mg/dL (N=7) spiked with hemolysate (HI range: 7-191) had an average difference of -0.1 mg/dL (range: -0.2-0.0 mg/dL) from non-hemolyzed pools, D-bil pools 0.4-0.9 mg/dL (N=10; HI range: 6-173) had an average difference of -0.2 mg/dL (range: -0.5-0.0 mg/dL), and D-bil pools >1.0 mg/dL (N=9; HI range: 5-182) had an average difference of -20.7% (range –45.5-0.0%). Serum pools with D-bil <0.3 mg/dL (N=23; HI range: 43-191) tested using the decreased volume function had an average difference –0.1 mg/dL (range: -0.2-0.0 mg/dL; HI range: 22-96) from the non-hemolyzed specimens, D-bil pools 0.4-0.9 mg/dL (N=20; HI range: 39-173) had an average difference -0.1 mg/dL (-0.3-0.0 mg/dL; HI range: 19-86), and D-bil pools >1.0 mg/dL (N=12; HI range: 37-182 ) had an average difference of -0.2 mg/dL (-0.5-0.2 mg/dL; HI range: 19-91). Conclusion We established a new hemolysis interference threshold of 75 for samples with D-bil <0.3 mg/dL. Using the decreased volume function, we expanded the hemolysis interference threshold to HI=100 for specimens =0.3 mg/dL. All specimens with HI>100 will result in a recollection. With these changes, we estimate that we can reduce our D-bil recollections by up to 6% over six months.
A-207 Investigating an Increase in Specimen Recollection Rates Due to Hemolysis After a Testing Location Change: The Impact of Specimen Type and Patient Population
Mohamed, Sulieman, Kelley, Brandon, Wockenfus, Amy M., McGinn, Alec, Lexvold, Nathaniel, Swartchick, Chelsea B. , Katzman, Brooke M.

Clinical Chemistry | October 2, 2025 | Volume: 71 | Issue: Supplement_1 | Pages: No page available

10.1093/clinchem/hvaf086.201 Get It

Abstract Background The Oncology Department at our institution requested faster turn-around times (TAT) for chemistry testing. Prior to their request, serum separator tubes (SST) were collected and tested in our core laboratory. To accommodate the request, we added plasma total bilirubin and aspartate aminotransferase (AST) to our stat laboratory test menu with collection in plasma separator tubes (PST). Upon implementation of this testing change, we observed an increase in specimen recollection rates due to hemolysis. We systematically evaluated laboratory testing variables that could have contributed to the increase: pneumatic tube transport path, differences in hemolysis measurements between chemistry analyzer models, and specimen types. Previous studies have demonstrated that lithium heparin plasma specimens are more susceptible to hemolysis than serum. Likewise, patients undergoing chemotherapy or radiation therapy may have fragile red blood cell (RBC) membranes, further complicating the issue. Our study*s objective was to assess the impact of changing from SST to PST collections. We conducted a 3-month pilot study to determine if we could reduce our recollection rates by collecting/testing Rapid Serum Tube (RST), while maintaining plasma TAT. Methods We compared recollection rates for specimens collected from the oncology patients and tested in our core laboratory (SST, May-June 2021), stat laboratory (PST, September-November 2022), pilot period in stat lab (RST, January-March 2023), and post-pilot period in stat lab (PST, May-July 2023). Data were extracted from our laboratory information system and recollection rates were calculated using Tableau (Salesforce, Inc, San Francisco, CA). Analysis was performed in Excel (Microsoft, Redmond, WA) using Student’s t-test, with p < 0.05 defined as statistically significant. Results SST specimens collected from oncology patients and tested in the core lab (May-June 2021) resulted in a recollection rate of 0.17% (4/2341). PST specimens sent to the stat lab (Sept-Nov 2022) had an increased recollection rate of 0.90% (55/6129, p-value < 0.001). During the pilot period when RST tubes were used (Jan-Mar 2023), our recollection rate decreased to 0.19% (11/5801, p-value 0.014). After the pilot period ended and PST tubes were again utilized for testing (May-July 2023), the recollection rate again increased to 0.34% (22/6500, p-value 0.019). Conclusion Rapid TAT of laboratory results within specialized-care settings, such as a cancer treatment clinic, is necessary to provide patients with the most effective and personalized treatment regimens. Hemolysis presents a challenge as it generates an analytical issue with the release of contents from RBCs that may spuriously affect the concentration of the analyte being measured, delaying result reporting due to the need for specimen recollection. The data from our study highlight the utility of rapid serum tubes. Not only did they achieve stat TAT that was observed with the use of PST, but they also decreased specimen recollection rates that mirror the low rates observed with the use of SST. Thus, this technology may be an excellent replacement for standard SST when specimen integrity is critical to rapidly reporting accurate test results.
A-173 Evaluation of Plateletworks and TEG 6S Platelet Mapping: Performance Characteristics and Differentiation of Healthy Donors from Patients on Dual Antiplatelet Therapy
Wockenfus, Amy M., Kelley, Brandon, Spears, Grant M. , Karon, Brad S. , Swan, Elena Ashikhmina , Katzman, Brooke M.

Clinical Chemistry | October 2, 2025 | Volume: 71 | Issue: Supplement_1 | Pages: No page available

10.1093/clinchem/hvaf086.168 Get It

Abstract Background Platelet function tests (PFT) help assess bleeding risk and guide appropriate hemostatic measures in patients on antiplatelet therapy undergoing emergent surgical interventions. Light transmission aggregometry (LTA, BioData Corporation), the gold standard for assessing platelet function, cannot be performed rapidly enough for emergent interventions. In this study, we compared two PFT, Plateletworks (Helena Laboratories) and TEG 6S (TEG, Haemonetics) adenosine diphosphate-(ADP) and arachidonic acid (AA)-induced platelet function, evaluating analytical imprecision and their ability to differentiate healthy controls from patients with acute coronary syndrome (ACS) on dual antiplatelet therapy. Methods Fresh specimens were collected from healthy donors (HD, N=19) and a patient group (PG, N=17) on both aspirin and clopidogrel taken within 24 hours of procedure. Percent aggregation was measured according to manufacturers’ instructions using Plateletworks and TEG, utilizing both AA and ADP and compared to LTA performed on a PAP-8E Platelet Aggregometer. Two separate venipuncture collections from HD were used to obtain repeat measures of Plateletworks and TEG to assess variability in platelet function due to pre-analytical variables. Using a two-way nested ANOVA model on data from HD, we estimated variability between replicates of a single sample (intra-assay imprecision) and between blood draws within a single patient (inter-assay imprecision), reported as coefficient of variation (CV%). Platelet function results were summarized as median (interquartile range [IQR; 25th-75th percentiles]) and compared between groups using nonparametric Wilcoxon rank-sum test. Receiver operating characteristic (ROC) analysis (using duplicate means from first blood draw) was performed. Results Plateletworks AA % aggregation, medians (IQR) were 96.3 (95.3-97.1) for HD and 48.9 (46.8-54.8) for PG (p<0.001); ADP % aggregation were 97.0 (94.5-98.2) and 37.1 (28.0-53.0) for HD and PG, respectively (p<0.001). TEG AA % aggregation was 97.8 (94.4-99.8) for HD and 13.5 (7.9-26.8) for PG (p<0.001); ADP % aggregation was 98.3 (95.3-99.9) for HD and 83.3 (77.8-87.7) for TG (p<0.001). LTA AA % aggregation was 55.0 (44.0-64.8) for HD and 0.0 (0.0-1.5) for PG (p<0.001); ADP % aggregation of 59.0 (40.8-70.8) and 14.5 (4.5-19.5) for HD and PG, respectively (p<0.001). Intra- and inter-assay imprecision estimates (CV%), respectively, in HD were 1.4 and 5.1 for Plateletworks AA, 5.5 and 15.9 for Plateletworks ADP, 3.0 and 0.0 for TEG AA, and 2.6 and 0.8 for TEG ADP. Intra-assay precision estimates for LTA AA and LTA ADP were 19.3 and 8.2. In PG, intra-assay precision was estimated as 7.0 for Plateletworks AA, 9.6 for Plateletworks ADP, 61.3 for LTA AA, 11.9 for LTA ADP, 37.6 for TEG AA, and 4.1 for TEG ADP. In our limited data set, ROC analysis showed optimal cutpoints (sensitivity/specificity) for Plateletworks AA 65.5 (1.0/1.0) and ADP 84.5 (1.0/0.90), TEG AA 87.1 (1.0/1.0) and ADP 92.7 (1.0/1.0), and LTA AA 8.75 (0.94/1.0) and ADP 25.0 (0.82/0.90) Conclusion Plateletworks most effectively distinguished % aggregation between healthy donors and patients on dual antiplatelet therapy. TEG also differentiated the groups, but differences in median ADP-induced aggregation were smaller than with Plateletworks. Both AA and ADP assays on LTA showed more variability than either Plateletworks or TEG.
Unintended Consequence of High-Dose Vitamin C Therapy for an Oncology Patient: Evaluation of Ascorbic Acid Interference With Three Hospital-Use Glucose Meters
Katzman, Brooke M. , Kelley, Brandon, Deobald, Gayle R., Myhre, Nikki K., Agger, Sean A., Karon, Brad S. , Kelley, Brandon R

Journal of Diabetes Science and Technology | June 7, 2020 | Volume: 15 | Issue: 4 | Pages: 897-900

10.1177/1932296820932186 Get It

The use of high-dose vitamin C in cancer care has offered promising results for some patients. However, the intravenous (IV) doses used for these patients can reach concentrations that interfere with some strip-based glucose meters. We characterized the impact of vitamin C interference, from standard to the very high doses used for some cancer protocols, using three different hospital-use glucose meters. For two of the three devices tested, increasing concentrations of ascorbic acid caused false elevations in the glucose measurements. The third glucose meter did not provide inaccurate results, regardless of the vitamin C concentration present. Rather, above a certain threshold, the device generated error messages and no results could be obtained.
Real-Time Detection of Analytical Outliers on the Roche Troponin T Generation 5 Assay
Wockenfus, Amy M., Hartung, Katherine J., Kelley, Brandon, Katzman, Brooke M. , Jaffe, Allan S. , Karon, Brad S. , Donato, Leslie J. , Jaffe, Allan S.

American Journal of Clinical Pathology | September 11, 2019 | Volume: 152 | Issue: Supplement_1 | Pages: S83-S83

10.1093/ajcp/aqz116.007 Get It

Abstract Introduction Analytical outliers occur with most troponin assays and can adversely affect patient management. We developed a system to detect analytical outliers for the Roche Troponin T (cTnT) Gen 5 assay using repeat analysis of samples manifesting changing values over 2 to 6 hours. Methods In our ED practice, troponins are collected at baseline, 2 hours, and at times at 6 hours in plasma separator tubes and processed on the Roche Cobas e411 (Indianapolis, IN). When troponin values change by ≥10 ng/L or ≤10 ng/L in the 2- or 6-hour sample compared to baseline, the sample that produced the result is repeated. If initial and repeat values differ by > ±5 ng/L (cTnT <100 ng/L) or ±5% (cTnT ≥100 ng/L), the initial value is deemed to be an analytical outlier. Outliers are confirmed by a third measurement. Results Since March 2018, 19 of 17,154 (0.11%) troponin samples analyzed have been labeled outliers. Outliers were falsely elevated in 14 cases and falsely decreased in 5. Mean (range) of initial cTnT values for outliers was 39 ng/L (18-621 ng/L), with only one initial value >100 ng/L. All changes could affect analysis of change over time. Eleven of 19 outliers were outside gender-specific reference interval on both samples and thus did not cause a change from abnormal to normal. Repeat measurements on 8 of 19 outliers were within the URL, which may have greater clinical consequences. Conclusions Repeat analysis upon observation of changing values allows detection of analytical outliers in real time, prior to reporting results. These outliers are almost always clinically significant in terms of interpretation of changing values and at times change patients from above to below the URL. If we tested more than just changing samples, we might detect still more outliers.