AbstractThe Fenwal Alyx for collecting double red cell products has two red cell volume collection settings: fixed collection target of 360 ml (180 ml/unit) and a variable target of collecting either 400 or 360 ml (200 or 180 ml/unit), where the machine aims for the higher possible collection target. We retrospectively compared the two collection targets for the RBC content, donor time, technician time, and collection efficiency. We compared 18 fixed (F) target collections to 40 variable (V) target collections. All collections were performed as per the manufacturer's recommendations on Alyx and donors met the manufacturer's eligibility criteria. There was no significant difference in average whole blood processed (F: 963 ml, V: 1,000 ml); donor time (F: 43 min, V: 45 min) or technician time (F: 64 min, V: 64 min). There was a significant difference in unit volume (F: 283 ml, V: 300 ml); grams Hb/unit (F: 53 g, V: 57 g); ml RBC/unit (F: 157 ml, V: 167 ml); and RBC recovery (F: 87.8%, V: 88.9%). The fixed target had a significantly lower frequency of products with ≥51 g Hb (80.6%) than variable target (96.3%) and ≥153 ml RBC/unit (F: 55.6%, V: 96.3%). In conclusion, the variable target efficiently allows collections of products with higher red cell volume and hemoglobin without a significant increase in collection and processing time.
Double red blood cell collection: comparison of three apheresis systems
BACKGROUNDWe compared three instruments for double red blood cell (DRBC) collection to efficiently optimize our RBC inventory.STUDY DESIGN AND METHODSInstruments compared were the TerumoBCT Trima Accel (Trima), the Fenwal Alyx (Alyx), and the Haemonetics MCS+ 8150 (MCS+). Forty consecutive collections (80 products) per instrument were evaluated for hemoglobin (Hb) content, leukoreduction, collection time, instrument efficiency, donor acceptance, and reactions. The total number of whole blood donors who could be eligible for DRBC collection was analyzed. All collections were as per the manufacturer recommendations with target volume of 360 or 400mL. Leukoreduction was integral to the collection procedure for the Alyx and Trima, while it was a separate process for the MCS+.RESULTSThe total numbers of whole blood donors who could be eligible for DRBC collection were 10,116 for the MCS+, 9378 for the Alyx, and 8573 for the Trima. All units collected had more than 42.5 g of Hb. Mean Hb levels were as follows: Trima, 59.2 g; Alyx, 56.8 g; and MCS+, 51.5 g. Donor times for the Trima, Alyx, and MCS+ were 52, 45, and 52 minutes, respectively. Technician times for the Trima, Alyx, and MCS+ (without filtration) were 87, 73, and 64 minutes. All collected products had fewer than 5 × 106 white blood cells.CONCLUSIONSAll three instruments would be capable of collecting an acceptable leukoreduced DRBC product. The Alyx was most portable, with the shortest donor time and total technician time (considering filtration time) and highest collection efficiency and would collect from more donors than the Trima. We thus chose the Alyx as it best fits the specific needs of our center.