RF Noise Artifacts From a High Frequency Spinal Cord Stimulation System: A Retrospective Imaging Cohort With Illustrative Cases
Stinson, Eric G. , Hubbard, Abby, Kleinow, Jenna E. , Kramer, Emily A., Trzasko, Joshua D. , Shu, Yunhong , Edmonson, Heidi A. , Hubbard, Abby J, Hubbard, Abby J.

Journal of Magnetic Resonance Imaging | July 6, 2026 | Volume: No volume available | Issue: No issue available | Pages: No page available

10.1002/jmri.70394 Get It

ABSTRACT Unexpected RF noise artifact “followed” a patient with an implanted spinal cord stimulator to multiple MR systems. Subsequent testing and clinical scanning showed no artifact, implying that the artifact originated from the implanted device. MR exams ( n = 173) performed over a two‐year period in patients ( n = 122) who had implanted spinal cord stimulators from the same manufacturer were reviewed for artifacts and device models. RF noise artifacts were observed in ~50% of exams in patients with a specific model of spinal cord stimulator, but only ~1%–8% of exams in other models from the same manufacturer. We hope that by describing this artifact, health care professionals can avoid service costs and scheduling delays.