Abstract Introduction Proximal perforation in complex penile prosthesis placement is well recognized. If recognized, suture sling repair amongst other techniques is our favored approach. Unrecognized proximal perforation often results in unilateral or bilateral perineal bruising, often with eventual proximal device migration and palpability of the proximal device in the buttock. Objective We present a series of 5 cases in which the suture sling technique was utilized for known proximal perforation, with subsequent need for immediate or delayed revision. Methods We assessed all cases performed at Mayo Clinic since 2018 that resulted in a known suture sling complication amongst 3 high volume implanters. We identified 5 cases in which the suture sling technique required revision surgery. All patients underwent cross sectional imaging for further evaluation. Results Patients were males aged 59 to 89 with erectile dysfunction. There were no consistent factors amongst the 5 cases that may have predicted suture sling complication. The majority of cases were inflatable penile prostheses (4/5), were prosthesis revision cases (3/5), utilized the penoscrotal approach (4/5), were Coloplast devices (4/5), utilized rear tip extenders (3/5), utilized Prolene suture sling (3/5), and returned to OR within 2 days for revision (3/5). The majority of the complications (4/5) involved excessive proximal cylinder length that were palpable on examination of the buttocks or near the rectum (one felt on rectal exam). One case demonstrated improper anchoring of the device not allowing proximal seating and thus S shaped deformity. We noted medial cylinder deviation on all imaging, as corporal perforation was always medial to the true corporal path. Conclusions When performing suture slings, surgeons should be very wary of oversizing. The medial cylinder path and prior surgical capsule may mislead surgeons when choosing appropriate cylinder length. If perforated, we advocate against rear tips and for under sizing in general, as proximal anchoring is reliant upon the suture, not a true corporal boundary. Similarly, the device should be fully inflated and proximally seated on palpation prior to tying down the sling to avoid S deformities. Disclosure No