A-301 Evaluation of Second-Generation Visby Medical Sexual Health Point-of-Care Test: Clinical Concordance, Analytical Error Rates, and Result Turnaround Time
Kelley, Barbara R, Wockenfus, Amy M., Binnicker, Matthew J. , Karon, Barry L., Kelley, B, Donato, Leslie J.

Clinical Chemistry | October 2, 2024 | Volume: 70 | Issue: Supplement_1 | Pages: No page available

10.1093/clinchem/hvae106.298 Get It

Abstract Background Rapid testing for Neisseria gonorrhoeae, Chlamydia trachomatis, and Trichomonas vaginalis may be useful for antibiotic stewardship initiatives. Our institution recently implemented the second-generation Visby Medical Sexual Health Test (Visby Gen2) for rapid diagnosis of STIs in our emergency department. Here we (1) compared result concordance of the (Visby Gen2) to the first-generation Visby Medical Sexual Health Click Test (Visby Gen1), (2) compared assay failure rates of the Visby Gen1 to Gen2 tests, and (3) compared result turnaround time (TAT) for clinically ordered rapid STI tests to central laboratory STI results. Methods Accuracy was assessed using residual clinical samples in Aptima buffer submitted for routine molecular STI testing (C. trachomatis, N. gonorrhoeae, and T. vaginalis) in the central laboratory on the Panther platform (Hologic). Aptima samples were diluted 1:33 in Visby collection buffer prior to testing on the Visby cartridge. Clinically reported results were used as reference results. Diluted samples with discordant results were re-tested on the reference method and the result from the dilution was used in calculations. Assay failure rates were calculated as the number of failed assays compared to the total assays performed. TAT for rapid and central lab testing was calculated as time of sample received in laboratory to time verified in laboratory information system. Results For Visby Gen1 we observed negative concordance of 100% (41/41) for C. trachomatis, 97.5% (39/40) for N. gonorrhoeae, and 97.1% (34/35) for T. vaginalis. Positive concordance was 91.7% (11/12) for C. trachomatis, 100% (11/11) for N. gonorrhoeae, and 100% (17/17) for T. vaginalis. For Visby Gen1 we observed a 10.5% (12/114) cartridge failure rate. For Visby Gen2, we observed negative concordance of 100% (59/59) for C. trachomatis, 100% (63/63) for N. gonorrhoeae, and 100% (55/55) for T. vaginalis while positive concordance was 100% (10/10) for C. trachomatis, 100% (6/6) for N. gonorrhoeae, and 90.9% (10/11) for T. vaginalis. The Visby Gen2 assay displayed a 4.9% (4/81) cartridge failure rate. Six weeks after implementing the Visby Gen2 assay in our ED, rapid STI panel TAT was 0.48 hours on average (n=20). In the central laboratory the average TAT was 22.31 hours (n=7892). Conclusions Assay concordance to reference method was similar for Visby Gen1 and Gen2 assays. The cartridge failure rate improved by 46.7% with the Gen2 assay which should decrease the need to repeat testing due to assay failure. Result TAT decreased by 97.8% when comparing rapid Visby results to central laboratory. Therefore, Visby STI testing using the Gen2 assay can provide accuracy and rapid results likely to facilitate more rapid patient management.