Patient doses in bone mineral densitometry
Huda, W, Morin, Richard L.

The British Journal of Radiology | July 15, 2009 | Volume: 69 | Issue: 821 | Pages: 422-425

10.1259/0007-1285-69-821-422 Get It

Abstract Procedures were developed to compute effective dose equivalent (HE) values for patients undergoing bone mineral densitometry (BMD) measurements such as dual energy X-ray absorptiometry (DEXA) and quantitative computed tomography (QCT). Representative values of HE were determined for patients undergoing each type of BMD procedure. Typical DEXA measurements are associated with an HE of about 2.5 μSv. For QCT, the values of HE are 300 μSv for single energy techniques and 1 mSv for dual energy techniques, respectively. By comparison, a single well collimated abdominal X-ray examination often taken in conjunction with DEXA studies results in an HE of about 100 μSv. BMD patient doses are at the lower end of the exposure range encountered in diagnostic radiology. As a result, radiation dose is not a primary factor in choosing the method for BMD determination.