Metal implants which saturate the CT number scale may require dosimetrist and physicist involvement to manually contour and assign an appropriate value to the metal for accurate dose calculation. This study investigated dose calculation based directly on extended CT scale images for different metals and geometries. The aim was to evaluate extended CT accuracy as a suitable alternative to standard CT methods in the presence of high‐Z materials and artifacts, despite the reduced HU resolution of extended CT. Gafchromic film measurements were made for comparison to calculated doses. The method of direct dose calculation on extended CT scale was compared to our institution's standard method of manually contouring and assigning metal values on saturated CT images for each of the metal samples. Clinical patient plans with metal implants were investigated and DVHs were compared between standard CT and extended CT dose calculations. Dose calculations showed agreement within 2% between the two methods of metal characterization and the film measurement in the case of the strongest metal attenuator, cobalt‐chromium. In the clinical treatment plans, the greatest dose discrepancy between the two methods was 1.2%. This study suggests that direct dose calculation on an extended scale CT image in the presence of metal implants can produce accurate clinically viable treatment plans, thereby improving efficiency of clinical workflow and eliminating a potential source of human error by manual CT number assignment.PACS number(s): 87.55.dk
Systematic offset of kV and MV localization systems as a function of gantry angle
Mechanical flex of the gantry and mounted imaging panels leads to systematic offsets in localization image isocenter as a function of gantry angle for linear accelerator‐mounted image guidance systems. Subsequently, object positions obtained from localization radiographs may be offset, resulting in greater target positioning uncertainty. While current QA procedures measure kV/MV image agreement, these measurements do not provide insight to apparent isocenter position for either single imaging system as a function of gantry rotation. This study measures offset as a function of gantry angle in kV and MV imaging systems on four treatment machines to investigate the magnitude of systematic offsets and their reproducibility between systems and machines, as well as over time. It is shown that each machine and energy has a reproducible pattern of offset as a function of gantry angle that is independent of kV/MV agreement, and it varies by machine. kV and MV offset ranges are on the order of 1.5 mm in the R/L and A/P directions, and 0.5 mm in the S/I direction. Variability of kV‐MV agreement is on the order of 0.7 mm. At certain angles, combinations of localization images could show a compounded offset of over 2 mm, exceeding the desired certainty threshold. Since these trends are persistent over time for each machine, online correction for image offsets as a function of gantry angle could improve the margin of positioning uncertainty.PACS number: 87.55.Qr