A Method to Improve the Accuracy of Ultrasound Geometrical Measurements of Boundaries in Soft Tissue
Pryor, A.T., Goldberg, Valentin , Parker, Dennis L.

Ultrasonic Imaging | March 2, 2013 | Volume: 9 | Issue: 2 | Pages: 117-131

10.1177/016173468700900204 Get It

The geometric distortion due to the neglect of refraction in existing ultrasound scanners is demonstrated. An iterative technique to correct for refraction errors in B-mode ultrasound, denoted Second Order Refraction Transposition (SORT), is presented. In vitro experiments were performed that confirm the calculated results.