Back to Authors

Wickremasinghe, Andrea C.

https://orcid.org/0000-0002-4228-9055

Mayo Clinic

Mayo Clinic

Citations 1

Accuracy of Neonatal Transcutaneous Bilirubin Measurement in the Outpatient Setting
Wickremasinghe, Andrea C. , Karon, Brad S. , Cook, Walter J.

Clinical Pediatrics | October 19, 2011 | Volume: 50 | Issue: 12 | Pages: 1144-1149

10.1177/0009922811417292 Get It

Objective. To evaluate the effectiveness of transcutaneous bilirubin (TcB) measurement in predicting risk for neonatal hyperbilirubinemia in outpatients. Design. Subjects were infants ≤8 days old seen in an outpatient clinic. Infants discharged with high-risk (HR) or high-intermediate risk (HIR) total serum bilirubin (TSB) values and jaundiced infants were recruited. TSB and TcB (BiliChek) levels were plotted on an hour-specific nomogram to determine risk for hyperbilirubinemia. Results. A total of 79 infants provided 87 sets of TcB and TsB values. Mean bias and standard deviation between TcB and TsB was 1.5 ± 2.1 mg/dL for outpatients, compared with 2.7 ± 1.3 mg/dL for inpatients. The sensitivity and specificity of HR or HIR TcB for predicting an HR or HIR TSB were 87% and 58%, respectively. Of 9 infants readmitted for phototherapy, 1 had a low-risk TcB and high-risk TSB. Conclusions. TcB screening in the outpatient environment may not be safe and efficient.