Precision across the Analytical Measuring Range of a Quantitative Real-Time PCR Assay for Cytomegalovirus Detection among Three Clinical Laboratories
Grys, Thomas E. , Duquette, Doreen L., White, Bruce A., Irish, Cole L., Hata, D. Jane , Pritt, Bobbi S. , Pritt, Bobbi S , Grys, Thomas E

Journal of Clinical Microbiology | June 9, 2011 | Volume: 49 | Issue: 8 | Pages: 3044-3046

10.1128/jcm.00486-11 Get It

ABSTRACT This study measured the precision of a quantitative laboratory-developed real-time PCR test for cytomegalovirus performed at three different clinical laboratories that use the same methodology. The overall standard deviation (adjusted for analyte level) was 0.18 log 10 copies/ml, and there was no significant relationship between standard deviation and analytical measuring range.