Abstract In Vitro testing is widely used to evaluate the hemodynamic performance of transcatheter aortic valves (TAVs). However, many studies assess self-expanding valves using blood analogs matched in viscosity at room temperature, despite the temperature-dependent mechanical behavior of nitinol stent frames. This study investigates the independent and combined effects of temperature and viscosity on the in vitro performance assessment of a self-expanding TAV. Rheological characterization was performed to identify water-glycerin solutions with matched viscosities at room temperature and physiological temperature (37 °C). A patient-specific aortic model was cast in silicone. A self-expanding Medtronic Evolut R valve was deployed within the compliant model and evaluated using a pulse-duplicating left heart simulator under physiological flow and pressure conditions. The transvalvular pressure gradient and effective orifice area (EOA) were calculated. Rheometry identified two fluid conditions with comparable viscosities at room and body temperatures. Heating the blood analog to 37 °C resulted in significantly reduced transvalvular pressure gradient and increased EOA compared to the viscosity-matched room temperature condition (p < 0.0001). At physiological body temperature, the higher viscosity fluid (42% glycerin) yielded further improvements in valve performance relative to the lower viscosity fluid (38% glycerin), with lower pressure drop and larger EOA (p < 0.0001). These results demonstrate that temperature-dependent effects influence the hemodynamic performance of self-expanding nitinol TAVs beyond viscosity matching alone. Combined consideration of temperature and viscosity is therefore necessary to ensure physiologically relevant in vitro assessment of self-expanding TAVs.
Residual leaks following percutaneous left atrial appendage occlusion and outcomes: a meta-analysis
Abstract Background and Aims Residual leaks are not infrequent after left atrial appendage occlusion. However, there is still uncertainty regarding their prognostic implications. The aim of this study is to evaluate the impact of residual leaks after left atrial appendage occlusion. Methods A literature search was conducted until 19 February 2023. Residual leaks comprised peri-device leaks (PDLs) on transoesophageal echocardiography (TEE) or computed tomography (CT), as well as left atrial appendage patency on CT. Random-effects meta-analyses were performed to assess the clinical impact of residual leaks. Results Overall 48 eligible studies (44 non-randomized/observational and 4 randomized studies) including 61 666 patients with atrial fibrillation who underwent left atrial appendage occlusion were analysed. Peri-device leak by TEE was present in 26.1% of patients. Computed tomography-based left atrial appendage patency and PDL were present in 54.9% and 57.3% of patients, respectively. Transoesophageal echocardiography-based PDL (i.e. any reported PDL regardless of its size) was significantly associated with a higher risk of thromboembolism [pooled odds ratio (pOR) 2.04, 95% confidence interval (CI): 1.52–2.74], all-cause mortality (pOR 1.16, 95% CI: 1.08–1.24), and major bleeding (pOR 1.12, 95% CI: 1.03–1.22), compared with no reported PDL. A positive graded association between PDL size and risk of thromboembolism was noted across TEE cut-offs. For any PDL of >0, >1, >3, and >5 mm, the pORs for thromboembolism were 1.82 (95% CI: 1.35–2.47), 2.13 (95% CI: 1.04–4.35), 4.14 (95% CI: 2.07–8.27), and 4.44 (95% CI: 2.09–9.43), respectively, compared with either no PDL or PDL smaller than each cut-off. Neither left atrial appendage patency, nor PDL by CT was associated with thromboembolism (pOR 1.45 and 1.04, 95% CI: 0.84–2.50 and 0.52–2.07, respectively). Conclusions Peri-device leak detected by TEE was associated with adverse events, primarily thromboembolism. Residual leaks detected by CT were more frequent but lacked prognostic significance.