Study Design: Prospective randomized Food and Drug Administration investigational device exemption clinical trial. Objective: The purpose of the present study is to report the 1-year clinical and radiographic outcomes and safety profile of patients who underwent lumbar facet arthroplasty through implantation of the Total Posterior Spine System (TOPS) device. Summary of Background Data: Lumbar facet arthroplasty is one proposed method of dynamic stabilization to treat grade-1 spondylolisthesis with stenosis; however, there are currently no Food and Drug Administration-approved devices for facet arthroplasty. Methods: Standard demographic information was collected for each patient. Radiographic parameters and patient-reported outcome measures were assessed preoperatively and at regular postoperative intervals. Complication and reoperation data were also collected for each patient. Results: At the time of this study, 153 patients had undergone implantation of the TOPS device. The mean surgical time was 187.8 minutes and the mean estimated blood loss was 205.7cc. The mean length of hospital stay was 3.0 days. Mean Oswestry Disability Index, Visual Analog Score leg and back, and Zurich Claudication Questionnaire scores improved significantly at all postoperative time points (P>0.001). There were no clinically significant changes in radiographic parameters, and all operative segments remained mobile at 1-year follow-up. Postoperative complications occurred in 11 patients out of the 153 patients (7.2%) who underwent implantation of the TOPS device. Nine patients (5.9%) underwent a total of 13 reoperations, 1 (0.6%) of which was for device-related failure owing to bilateral L5 pedicle screw loosening. Conclusions: Lumbar facet arthroplasty with the TOPS device demonstrated a statistically significant improvement in all patient-reported outcome measures and the ability to maintain motion at the index level while limiting sagittal translation with a low complication rate.
Catastrophic acute failure of pelvic fixation in adult spinal deformity requiring revision surgery: a multicenter review of incidence, failure mechanisms, and risk factors
OBJECTIVE There are few prior reports of acute pelvic instrumentation failure in spinal deformity surgery. The objective of this study was to determine if a previously identified mechanism and rate of pelvic fixation failure were present across multiple institutions, and to determine risk factors for these types of failures. METHODS Thirteen academic medical centers performed a retrospective review of 18 months of consecutive adult spinal fusions extending 3 or more levels, which included new pelvic screws at the time of surgery. Acute pelvic fixation failure was defined as occurring within 6 months of the index surgery and requiring surgical revision. RESULTS Failure occurred in 37 (5%) of 779 cases and consisted of either slippage of the rods or displacement of the set screws from the screw tulip head (17 cases), screw shaft fracture (9 cases), screw loosening (9 cases), and/or resultant kyphotic fracture of the sacrum (6 cases). Revision strategies involved new pelvic fixation and/or multiple rod constructs. Six patients (16%) who underwent revision with fewer than 4 rods to the pelvis sustained a second acute failure, but no secondary failures occurred when at least 4 rods were used. In the univariate analysis, the magnitude of surgical correction was higher in the failure cohort (higher preoperative T1-pelvic angle [T1PA], presence of a 3-column osteotomy; p < 0.05). Uncorrected postoperative deformity increased failure risk (pelvic incidence–lumbar lordosis mismatch > 10°, higher postoperative T1PA; p < 0.05). Use of pelvic screws less than 8.5 mm in diameter also increased the likelihood of failure (p < 0.05). In the multivariate analysis, a larger preoperative global deformity as measured by T1PA was associated with failure, male patients were more likely to experience failure than female patients, and there was a strong association with implant manufacturer (p < 0.05). Anterior column support with an L5–S1 interbody fusion was protective against failure (p < 0.05). CONCLUSIONS Acute catastrophic failures involved large-magnitude surgical corrections and likely resulted from high mechanical strain on the pelvic instrumentation. Patients with large corrections may benefit from anterior structural support placed at the most caudal motion segment and multiple rods connecting to more than 2 pelvic fixation points. If failure occurs, salvage with a minimum of 4 rods and 4 pelvic fixation points can be successful.