Percutaneous transendocardial injections of therapeutic agents into the myocardium may not always be effective. We used an animal model for assessing the efficacy of the injections using linoleic acid as a testing agent. Efficacious delivery into the myocardium of a beating heart was indicated by rapidly developed local myocardial necrosis and wall motion abnormalities using echocardiography. We employed this experimental model to test our innovative technology, an acoustically active injection catheter. The Doppler ultrasound‐guided acoustically active injection catheter effectively delivers the substance to the myocardium but needs further technical improvements to minimize an unwanted systemic distribution of the agent.
Acoustic navigation of intramyocardial injection needle catheter using color doppler echocardiography
B‐mode ultrasound imaging guidance of cannulas can be compromised by noise, artifacts, and echogenicity that is not distinctive from that of surrounding anatomy. We have modified a venovenous extracorporeal membrane oxygenation cannula by embedding piezoelectric crystals into each of its 3 blood flow ports. Each vibrating crystal acoustically interacts with a Doppler imaging signal and produces an instantaneous color marker. The aim of this study was to compare identification of the extracorporeal membrane oxygenation cannula ports by B‐mode imaging versus the color Doppler marker. Unlike B‐mode imaging, the color Doppler marker identified the corresponding port even in highly challenging closed‐chest scans in anesthetized pigs. The method could improve guidance accuracy of cannulas by ultrasound scans.