Abstract Purpose Although the Woven EndoBridge (WEB) device is increasingly used for the treatment of wide-neck intracranial aneurysms, including in the acute rupture setting, comparative evidence assessing the impact of rupture status remains limited. This study compared angiographic, safety, and clinical outcomes between ruptured and unruptured intracranial aneurysms treated with WEB. Methods We conducted a retrospective analysis of prospectively collected data from the multicenter cohort registry WorldWideWEB, including consecutive adult patients with intracranial aneurysms treated with the WEB. Patients were stratified into groups of ruptured and unruptured aneurysms. Propensity score matching was used to balance baseline characteristics between both groups. Retreatment rate was the primary outcome. Secondary outcomes included mRS, safety events (thromboembolic complications) and angiographic outcomes (periprocedurally and last follow-up). Results Among 1,220 patients, 342 (28.0%) presented with ruptured aneurysms. There was no significant difference in the rate of thromboembolic complications between groups (11.8% vs 5.9%, p = 0.056). At follow-up, adequate occlusion was lower (82.2% vs 93.3%, p < 0.002) and functional outcomes were worse (mRS ≥2 in 34.1% vs 21.9%, p = 0.012) among patients with ruptured aneurysms. Periprocedural adequate occlusion (53.3% vs 53.8%, p>0.9) and retreatment rates (11.8% vs 7.1%, p = 0.14) did not differ between groups after matching. Conclusion Ruptured aneurysms demonstrated expected inferior follow-up functional and angiographic outcomes when compared with unruptured aneurysms, but no difference in retreatment rate and procedural safety. These findings support WEB as a safe and effective treatment option for appropriately selected ruptured intracranial aneurysms in routine clinical practice.
Higher risk of recurrence in partially thrombosed cerebral aneurysms post-WEB (Woven EndoBridge) device treatment: insights from the WorldWideWEB Consortium registry
Journal of NeuroInterventional Surgery | April 30, 2025 | Volume: 18 | Issue: 3 | Pages: 782-789
Background The Woven EndoBridge (WEB) device is a prevalent treatment for intracranial aneurysms. While many studies have assessed the obliteration rate post-WEB embolization, few have focused on long-term outcomes in partially thrombosed aneurysms. Objective To assess whether partially thrombosed aneurysms are at higher risk of recurrence or retreatment following WEB embolization compared with non-thrombosed aneurysms. Methods We evaluated data from 22 academic institutions, focusing on previously untreated cerebral aneurysms treated with the WEB device. Logistic regression was utilized to analyze factors predicting long-term aneurysm obliteration and retreatment necessity. Results Among 1303 patients, 26 presented with a partially thrombosed aneurysm. In the partially thrombosed group, the mean aneurysm maximal diameter was 10.7±4 mm with a neck ratio of 1.99±1.19 mm, larger than in the control group where the mean aneurysm maximal diameter was 6.81±2.37 mm with a neck ratio of 1.64±0.51 mm (P<0.001 for both maximal diameter and neck ratio). At the final follow-up, partially thrombosed aneurysms treated by the WEB device had a 38.5% retreatment rate, compared with 7.0% for non-thrombosed aneurysms (P<0.001). Among partially thrombosed aneurysms, the Raymond-Roy type IIIa/b occlusion rate was higher (38.5% vs 9.9%, P<0.001). On multivariate analysis, partially thrombosed aneurysms compared with non-thrombosed aneurysms had an increased rate of retreatment (OR 3.64, 95% CI 1.28 to 10.1). Conclusion Partially thrombosed aneurysms are associated with a poorer occlusion rate and a higher rate of retreatment following WEB embolization. For partially thrombosed aneurysms, the WEB device appears suboptimal as a first-line treatment, and therefore alternative techniques should be prioritized.