A-008 Comparison of Roche 5th Generation Troponin T Analytical Outlier Rates Between the e411 Analyzer Using a Biotin-sensitive Reagent and the e601 Analyzer Using Reformulated (Biotin-insensitive) Reagent
Clinical Chemistry | September 27, 2023 | Volume: 69 | Issue: Supplement_1 | Pages: No page available
Abstract Background Analytical outliers (fliers) occur with most troponin methods and can adversely affect patient management. We compared the analytical outlier rates for the Roche 5th generation Troponin T STAT (cTnT) reformulated (biotin-insensitive) reagent using a Cobas e601 analyzer (Indianapolis, IN) to rates observed in a previous study using the biotin-sensitive reagent on a Cobas e411 analyzer. Methods Plasma separator tubes (PST) were collected from hospital patients with orders for clinical cTnT testing. PST samples were centrifuged for 3 min at 4000g and analyzed using the biotin-insensitive cTnT Gen 5 assay on a Roche Cobas e601 immunoassay system. If samples displayed measurable cTnT Gen 5 (≥6 ng/L), samples were stored at 2–8 °C within 2 h of initial measurement (N = 2555). Within 24 h of initial analysis, samples were warmed to room temperature, aliquoted, re-centrifuged at 4000g for 3 min, and cTNT was re-measured on the same analyzer. Outlier rates were compared to a previous study (N = 1997) using nearly identical study design but using biotin-sensitive Roche 5th Gen cTnT reagent on an e411 analyzer. We defined analytical outliers by three methods: Repeat value (after aliquoting and re-centrifugation) that differed from the initial value for all samples with measurable cTnT by a critical difference (CD): where CD = z × √2 × SDanalytical, with a z-value of 3.719 and using SD from best fit line of precision vs cTnT concentration. Precision experiments were performed using four plasma pools in 5 × 5 precision experiment on the e411 analyzer. Using the same CD but considering only initial cTnT values <100 ng/L, those are most likely to be clinically significant or lead to misinterpretation. Using CD for initial values <100 ng/L, and a fixed 10% difference between replicates for initial values ≥100 ng/L, accounting for our practice of reporting the cTnT “delta” in percent (rather than absolute) for values ≥100 ng/L. Comparisons of outlier rates between the biotin-sensitive and biotin-insensitive reagents were performed using Pearson’s chi-squared test. Results Using the CD the total number of outliers that were observed for the biotin-insensitive reagent on the e601 was 208/2555 (8.1%), compared to 68/1997 (3.4%) on the biotin-sensitive reagent measured on the e411 (P < 0.0001). Outliers were observed in 138 (6.4%) of samples with initial values < 100 ng/L using the biotin-insensitive reagent on the e601 compared to 45 (2.6%) of such samples on the e411 (P < 0.0001). Using the Z score method to define outliers for initial values <100 ng/L and a fixed 10% to define outliers for initial values ≥100 ng/L, the outlier rate was 151 (5.9%) for the biotin-insensitive reagent on the e601 compared to 53 (2.7%) on the e411 (P < 0.0001). Conclusions Analytical outliers occurred more frequently using the reformulated (biotin-insensitive) reagent on the Roche Cobas e601 analyzer, compared to outlier rates observed for biotin-sensitive reagent on the Cobas e411.
Cardiac troponin measurement at the point of care: educational recommendations on analytical and clinical aspects by the IFCC Committee on Clinical Applications of Cardiac Bio-Markers (IFCC C-CB)
Clinical Chemistry and Laboratory Medicine (CCLM) | January 13, 2023 | Volume: 61 | Issue: 6 | Pages: 989-998
Abstract The International Federation of Clinical Chemistry and Laboarator Medicine (IFCC) Committee on Clinical Applications of Cardiac Bio-Markers (C-CB) has provided evidence-based educational resources to aid and improve the understanding of important analytical and clinical aspects of cardiac biomarkers. The present IFCC C-CB educational report focuses on recommendations for appropriate use, analytical performance, and gaps in clinical studies related to the use of cardiac troponin (cTn) by point of care (POC) measurement, often referred to as a point of care testing (POCT). The use of high-sensitivity (hs)-cTn POC devices in accelerated diagnostic protocols used in emergency departments or outpatient clinics investigating acute coronary syndrome has the potential for improved efficacy, reduction of length of stay and reduced costs in the health care system. POCT workflow integration includes location of the instrument, assignment of collection and testing responsibility to (non-lab) staff, instrument maintenance, in-service and recurrent training, quality control, proficiency assessments, discrepant result trapping, and troubleshooting and inventory management.
AACC Guidance Document on the Use of Point-of-Care Testing in Fertility and Reproduction
The Journal of Applied Laboratory Medicine | September 1, 2022 | Volume: 7 | Issue: 5 | Pages: 1202-1236
Abstract Background The AACC Academy revised the reproductive testing section of the Laboratory Medicine Practice Guidelines: Evidence-Based Practice for Point-of-Care Testing (POCT) published in 2007. Methods A panel of Academy members with expertise in POCT and laboratory medicine was formed to develop guidance for the use of POCT in reproductive health, specifically ovulation, pregnancy, premature rupture of membranes (PROM), and high-risk deliveries. The committee was supplemented with clinicians having Emergency Medicine and Obstetrics/Gynecology training. Results Key recommendations include the following. First, urine luteinizing hormone (LH) tests are accurate and reliable predictors of ovulation. Studies have shown that the use of ovulation predicting kits may improve the likelihood of conception among healthy fertile women seeking pregnancy. Urinary LH point-of-care testing demonstrates a comparable performance among other ovulation monitoring methods for timing intrauterine insemination and confirming sufficient ovulation induction before oocyte retrieval during in vitro fertilization. Second, pregnancy POCT should be considered in clinical situations where rapid diagnosis of pregnancy is needed for treatment decisions, and laboratory analysis cannot meet the required turnaround time. Third, PROM testing using commercial kits alone is not recommended without clinical signs of rupture of membranes, such as leakage of amniotic fluid from the cervical opening. Finally, fetal scalp lactate is used more than fetal scalp pH for fetal acidosis due to higher success rate and low volume of sample required. Conclusions This revision of the AACC Academy POCT guidelines provides recommendations for best practice use of POCT in fertility and reproduction.
A Contemporary Analysis of Phlebotomy and Iatrogenic Anemia Development Throughout Hospitalization in Critically Ill Adults
BACKGROUND: Anemia is common in critically ill patients and may be exacerbated through phlebotomy-associated iatrogenic blood loss. Differences in phlebotomy practice across patient demographic characteristics, clinical features, and practice environments are unclear. This investigation provides a comprehensive description of contemporary phlebotomy practices for critically ill adults. METHODS: This is an observational cohort study of adults ≥18 years of age requiring intensive care unit (ICU) admission between January 1, 2019, and December 31, 2019, at a large academic medical center. Descriptive statistics were utilized to summarize all phlebotomy episodes throughout hospitalization, with each phlebotomy episode defined by unique peripheral venous, central venous, or arterial accesses for laboratory draws, exclusive of finger sticks. Secondarily, financial costs of phlebotomy and the relationships between phlebotomy practices, hemoglobin concentrations, and red blood cell (RBC) transfusions were evaluated. RESULTS: A total of 6194 patients were included: 59% were men with a median (interquartile range) age of 66 (54–76) years and median ICU and hospital durations of 2.1 (1.4–3.9) and 7.1 (4.3–11.8) days, respectively. The median number of unique laboratory draws was 41 (18–88) throughout hospitalization, with a median volume of 232 (121–442) mL, corresponding to 5.2 (2.6–8.8) draws and 29 (19–43) mL per day. Waste (ie, discard) volume was responsible for 10.8% of total phlebotomy volume. Surgical patients had a higher number of phlebotomy episodes and greater total phlebotomy volumes compared to nonsurgical patients. Phlebotomy practices differed across ICU types, with the greatest frequency of laboratory draws in the cardiac surgical ICU and the greatest daily phlebotomy volume in the medical ICU. Across hospitalization, ICU environments had the greatest frequency and volumes of laboratory draws, with the least intensive phlebotomy practice observed in the general hospital wards. Patients in the highest quartile of cumulative blood drawn experienced the longest hospitalizations, lowest nadir hemoglobin concentrations, and greatest RBC transfusion utilization. Differences in phlebotomy practice were limited across patient age, gender, and race. Hemoglobin concentrations declined during hospitalization, congruent with intensity of phlebotomy practice. Each 100 mL of phlebotomy volume during hospitalization was associated with a 1.15 (95% confidence interval [CI], 1.14–1.17; P < .001) multiplicative increase in RBC units transfused in adjusted analyses. Estimated annual phlebotomy costs exceeded $15 million (approximately $2500 per patient admission). CONCLUSIONS: Phlebotomy continues to be a major source of blood loss in hospitalized patients with critical illness, and more intensive phlebotomy practices are associated with lower hemoglobin concentrations and greater transfusion utilization.
Analytical Sensitivity and Specificity of Four Point of Care Rapid Antigen Diagnostic Tests for SARS-CoV-2 Using Real-Time Quantitative PCR, Quantitative Droplet Digital PCR, and a Mass Spectrometric Antigen Assay as Comparator Methods
Abstract Background We evaluated the analytical sensitivity and specificity of 4 rapid antigen diagnostic tests (Ag RDTs) for severe acute respiratory syndrome coronavirus 2, using reverse transcription quantitative PCR (RT-qPCR) as the reference method and further characterizing samples using droplet digital quantitative PCR (ddPCR) and a mass spectrometric antigen test. Methods Three hundred fifty (150 negative and 200 RT-qPCR positive) residual PBS samples were tested for antigen using the BD Veritor lateral flow (LF), ACON LF, ACON fluorescence immunoassay (FIA), and LumiraDx FIA. ddPCR was performed on RT-qPCR–positive samples to quantitate the viral load in copies/mL applied to each Ag RDT. Mass spectrometric antigen testing was performed on PBS samples to obtain a set of RT-qPCR–positive, antigen-positive samples for further analysis. Results All Ag RDTs had nearly 100% specificity compared to RT-qPCR. Overall analytical sensitivity varied from 66.5% to 88.3%. All methods detected antigen in samples with viral load >1 500 000 copies/mL RNA, and detected ≥75% of samples with viral load of 500 000 to 1 500 000 copies/mL. The BD Veritor LF detected only 25% of samples with viral load between 50 000 to 500 000 copies/mL, compared to 75% for the ACON LF device and >80% for LumiraDx and ACON FIA. The ACON FIA detected significantly more samples with viral load <50 000 copies/mL compared to the BD Veritor. Among samples with detectable antigen and viral load <50 000 copies/mL, sensitivity of the Ag RDT varied between 13.0% (BD Veritor) and 78.3% (ACON FIA). Conclusions Ag RDTs differ significantly in analytical sensitivity, particularly at viral load <500 000 copies/mL.
Clarity on the Use of Glucose Meters for Critically Ill Hospitalized Patients, But One Big Question Remains to be Answered
The Journal of Applied Laboratory Medicine | May 1, 2021 | Volume: 6 | Issue: 4 | Pages: 813-815
Abstract Objectives To evaluate the analytical and clinical performance characteristics of the fifth-generation troponin T reagent. Methods Troponin T was measured in 2,332 paired serum and plasma samples from emergency department and hospital patients using the fourth- and fifth-generation reagents. Testing was repeated after recentrifugation to determine the frequency of analytical outliers and percentage of patients with elevated values for each assay. We conducted separate experiments to determine the effects of biotin and hemolysis interference, as well as measure interinstrument variability, for fifth-generation troponin T. Results Analytic outliers occurred more frequently using the fifth-generation reagent (3.4%) compared with the fourth-generation reagent (1.0%). The frequency of elevated troponin T above the 99th percentile upper reference limit was 26% for the fourth-generation reagent and 52% for the fifth-generation reagent. Clinically significant assay interference by biotin was observed at 20 ng/mL, but hemolysis interference was not observed until an H index of 150. Instrument-to-instrument variability between e411 and e601/602 instrument platforms is predicted to confound clinical interpretation of troponin changes. Conclusions Analytical outliers and instrument-to-instrument variability are the two analytical variables most likely to confound interpretation of changes in fifth-generation troponin T results over time.
Leadership and Management Training for Residents and Fellows: A Curriculum for Future Medical Directors
Archives of Pathology & Laboratory Medicine | February 11, 2021 | Volume: 131 | Issue: 4 | Pages: 610-614
Abstract Context. —Management of laboratories and pathology practices is increasingly complex. Residents and fellows in laboratory medicine and pathology need more structured curricula in leadership and management (L&M) training to function as medical and laboratory directors. Objective. —To define a curriculum that provides basic competency in L&M for residents and fellows in pathology. Design. —A year-long formal L&M course included didactic lectures, interactive sessions, case scenarios, team-building exercises, formal team presentations (capstone project), and precourse and postcourse assessment of L&M knowledge. The curriculum meets requirements of American College of Graduate Medical Education and supports goals for leadership training of the College of American Pathologists. Participants evaluated (5-point scale) the content and speakers of all sessions. Trainees were evaluated after considering postcourse examination results, quality of the capstone presentation, and a global assessment. Results. —The 5 non-capstone sessions received evaluation scores ranging from 4.4 (informatics) to 5 (L&M basics). Postcourse test scores showed significant improvement when compared with the pretest scores for the 2003– 2004 and 2004–2005 trainee cohorts. Conclusions. —Short-term results indicate that the course described improves trainee knowledge of L&M issues.
Error simulation modeling to assess the effects of bias and precision on bilirubin measurements used to screen for neonatal hyperbilirubinemia
Clinical Chemistry and Laboratory Medicine (CCLM) | January 23, 2021 | Volume: 59 | Issue: 6 | Pages: 1069-1075
Abstract Objectives Error simulation models have been used to understand the relationship between analytical performance and clinical outcomes. We developed an error simulation model to understand the effects of method bias and precision on misclassification rate for neonatal hyperbilirubinemia using an age-adjusted risk assessment tool. Methods For each of 176 measured total bilirubin (TSB M ) values, 10,000 simulated total bilirubin (TB S ) values were generated at each combination of bias and precision conditions for coefficient of variation (CV) between 1 and 15%, and for biases between −51.3 μmol/L and 51.3 μmol/L (−3 and 3 mg/dL) fixed bias. TB S values were analyzed to determine if they were in the same risk zone as the TSB M value. We then calculated sensitivity and specificity for prediction of ≥75th percentile for postnatal age values as a function of assay bias and precision, and determined the rate of critical errors (≥95th percentile for age TSB M with <75th percentile TB S ). Results A sensitivity >95% for predicting ≥75th percentile bilirubin values was observed when there is a positive fixed bias of greater than 17.1 μmol/L (1.0 mg/dL) and CV is maintained ≤10%. A specificity >70% for predicting <75th percentile bilirubin values was observed when positive systematic bias was 17.1 μmol/L (1 mg/dL) or less at CV ≤ 10%. Critical errors did not occur with a frequency >0.2% until negative bias was −17.1 μmol/L (−1 mg/dL) or lower. Conclusions A positive systematic bias of 17.1 μmol/L (1 mg/dL) may be optimal for balancing sensitivity and specificity for predicting ≥75th percentile TSB values. Negative systematic bias should be avoided to allow detection of high risk infants and avoid critical classification errors.
Test Utilization Proposal for Reflex Bilirubin Testing: Why Order Two Tests When One Will Do?
The Journal of Applied Laboratory Medicine | November 25, 2020 | Volume: 6 | Issue: 4 | Pages: 980-984
Abstract Background Reflex testing algorithms are effective tools to reduce unnecessary laboratory testing. Direct (conjugated) bilirubin (DB) and total bilirubin (TB) are often ordered together at our institution. Therefore, the objective of our study was to evaluate the potential impact of performing reflex testing for DB when TB is elevated. Methods We performed a retrospective review of test orders (patients ≥18 years of age) for DB, TB, or for both DB and TB on the same accession number received in our stat laboratory from January through April 2017. The orders were binned into 4 categories depending on the results from each individual test: (a) DB normal and TB normal, (b) DB normal and TB high, (c) DB high and TB normal, and (d) DB high and TB high. The percentage of orders and median (range) test result for each category was calculated. Results During the months evaluated, a total of 4828 stat orders were placed for DB, TB, or both DB and TB. A total of 4296 stat orders (89%) were placed with both DB and TB on the same accession number for 4158 unique patients. Of those orders, the vast majority of tests (87.3%) contained normal results for both analytes; only 12.7% of orders contained ≥1 abnormal result. Conclusions The majority of all bilirubin tests ordered stat for emergency department and hospitalized patients have values within the reference interval. Consequently, if reflex testing were executed on elevated TB, a large number of DB tests could be avoided.