Background: Closed-loop control (CLC) has been shown to improve glucose time in range and other glucose metrics; however, randomized trials >3 months comparing CLC with sensor-augmented pump (SAP) therapy are limited. We recently reported glucose control outcomes from the 6-month international Diabetes Closed-Loop (iDCL) trial; we now report patient-reported outcomes (PROs) in this iDCL trial. Methods: Participants were randomized 2:1 to CLC ( N = 112) versus SAP ( N = 56) and completed questionnaires, including Hypoglycemia Fear Survey, Diabetes Distress Scale (DDS), Hypoglycemia Awareness, Hypoglycemia Confidence, Hyperglycemia Avoidance, and Positive Expectancies of CLC (INSPIRE) at baseline, 3, and 6 months. CLC participants also completed Diabetes Technology Expectations and Acceptance and System Usability Scale (SUS). Results: The Hypoglycemia Fear Survey Behavior subscale improved significantly after 6 months of CLC compared with SAP. DDS did not differ except for powerless subscale scores, which worsened at 3 months in SAP. Whereas Hypoglycemia Awareness and Hyperglycemia Avoidance did not differ between groups, CLC participants showed a tendency toward improved confidence in managing hypoglycemia. The INSPIRE questionnaire showed favorable scores in the CLC group for teens and parents, with a similar trend for adults. At baseline and 6 months, CLC participants had high positive expectations for the device with Diabetes Technology Acceptance and SUS showing high benefit and low burden scores. Conclusion: CLC improved some PROs compared with SAP. Participants reported high benefit and low burden with CLC. Clinical Trial Identifier: NCT03563313.
Clinical Management and Pump Parameter Adjustment of the Control-IQ Closed-Loop Control System: Results from a 6-Month, Multicenter, Randomized Clinical Trial
Background: Data are limited on the need for and benefits of pump setting optimization with automated insulin delivery. We examined clinical management of a closed-loop control (CLC) system and its relationship to glycemic outcomes. Materials and Methods: We analyzed personal parameter adjustments in 168 participants in a 6-month multicenter trial of CLC with Control-IQ versus sensor-augmented pump (SAP) therapy. Preset parameters (BR = basal rates, CF = correction factors, CR = carbohydrate ratios) were optimized at randomization, 2 and 13 weeks, for safety issues, participant concerns, or initiation by participants' usual diabetes care team. Time in range (TIR 70–180 mg/dL) was compared in the week before and after parameter changes. Results: In 607 encounters for parameter changes, there were fewer adjustments for CLC than SAP (3.4 vs. 4.1/participant). Adjustments involved BR (CLC 69%, SAP 80%), CR (CLC 68%, SAP 50%), CF (CLC 44%, SAP 41%), and overnight parameters (CLC 62%, SAP 75%). TIR before and after adjustments was 71.2% and 71.3% for CLC and 61.0% and 62.9% for SAP. The highest baseline HbA 1c CLC subgroup had the largest TIR improvement (51.2% vs. 57.7%). When a CR was made more aggressive in the CLC group, postprandial time >180 mg/dL was 43.1% before the change and 36.0% after the change. The median postprandial time <70 mg/dL before making CR less aggressive was 1.8%, and after the change was 0.7%. Conclusions: No difference in TIR was detected with parameter changes overall, but they may have an effect in higher HbA 1c subgroups or following user-directed boluses, suggesting that changes may matter more in suboptimal control or during discrete periods of the day. Clinical Trials Registration number: NCT03563313.