Evaluation of LightCycler PCR for Implementation of Laboratory Diagnosis of Herpes Simplex Virus Infections

Journal of Clinical Microbiology | December 31, 2019 | Volume: 38 | Issue: 8 | Pages: 3116-3118

10.1128/jcm.38.8.3116-3118.2000 | PMID 10921993

ABSTRACT Five hundred specimens (288 genital, 192 dermal, and 20 ocular) were extracted by technologists, and the DNA was assayed by LightCycler PCR (DNA polymerase and thymidine kinase [TK] gene targets) and by conventional tube and shell vial cell culture. One hundred fifty-eight confirmed (by cell culture and TK target PCR) positive and LightCycler-positive specimens were detected during the first 30 PCR cycles. LightCycler PCR-positive results for cycles 31 to 45 (39 of 67 [58.2%]) required confirmation by another PCR target (TK). LightCycler PCR is more sensitive ( n = 197; 23.1%) than cell cultures ( n = 150) for the routine laboratory detection of herpes simplex virus infections.