Technical Note: An efficient daily QA procedure for proton pencil beam scanning
Younkin, James E. , Shen, Jiajian , Robertson, Daniel , Mundy, Daniel W. , Clouser, Edward , Ding, Xiaoning, Stoker, Joshua B., Bues, Martin , Robertson, Daniel G, Shen, Jiajian , Liu, Wei , Clouser, Edward Lee

Medical Physics | February 2, 2018 | Volume: 45 | Issue: 3 | Pages: 1040-1049

10.1002/mp.12787 Get It

PurposeThe aim of this work was to develop an efficient daily quality assurance (QA) program with strict tolerance levels for pencil beam scanning (PBS) proton radiotherapy featuring simultaneous dosimetric testing on a single, nonuniform field.MethodsA nonuniform field measuring beam output, proton range, and spot position was designed for delivery onto a Sun Nuclear Daily‐QA 3 device. A custom acrylic block permitted simultaneous measurement of low‐ and high‐energy proton ranges in addition to beam output. Sensitivities to output, range, and spot position were evaluated to quantitate the device's response. Reproducibility tests were used to identify and control sources of measurement error as well as to assess the QA procedure's robustness. This procedure was implemented in each of our four treatment rooms independently; 4–6 months of daily QA measurements were collected.ResultsThe 1% output, 0.5 mm range, and 1.5 mm spot position tolerances derived from preliminary tests were tighter overall than tolerances found in the literature and equivalent to the limits used for proton system commissioning. The simplicity and automation of the procedure reduced the time required for daily QA to 10 min per treatment room, and competition for beam between multiple treatment rooms was minimized.ConclusionsAn efficient daily PBS QA procedure can be performed using a single, nonuniform field on a nondedicated QA device. A thorough quantitation of the device's response and careful control of measurement uncertainties allowed daily tolerances to match commissioning standards.
A novel and fast method for proton range verification using a step wedge and 2D scintillator
Shen, Jiajian , Allred, Bryce C., Robertson, Daniel , Remmes, Nicholas B. , Keole, Sameer R., Bues, Martin , Robertson, Daniel G, Liu, Wei , Sio, Terence T. , Shen, Jiajian , Bues, Martin

Medical Physics | June 30, 2017 | Volume: 44 | Issue: 9 | Pages: 4409-4414

10.1002/mp.12439 Get It

PurposeTo implement and evaluate a novel and fast method for proton range verification by using a planar scintillator and step wedge.MethodsA homogenous proton pencil beam plan with 35 energies was designed and delivered to a 2D flat scintillator with a step wedge. The measurement was repeated 15 times (3 different days, 5 times per day). The scintillator image was smoothed, the Bragg peak and distal fall off regions were fitted by an analytical equation, and the proton range was calculated using simple trigonometry. The accuracy of this method was verified by comparing the measured ranges to those obtained using an ionization chamber and a scanning water tank, the gold standard. The reproducibility was evaluated by comparing the ranges over 15 repeated measurements. The sensitivity was evaluated by delivering to same beam to the system with a film inserted under the wedge.ResultsThe range accuracy of all 35 proton energies measured over 3 days was within 0.2 mm. The reproducibility in 15 repeated measurements for all 35 proton ranges was ±0.045 mm. The sensitivity to range variation is 0.1 mm for the worst case. This efficient procedure permits measurement of 35 proton ranges in less than 3 min. The automated data processing produces results immediately. The setup of this system took less than 5 min. The time saving by this new method is about two orders of magnitude when compared with the time for water tank range measurements.ConclusionsA novel method using a scintillator with a step wedge to measure the proton range was implemented and evaluated. This novel method is fast and sensitive, and the proton range measured by this method was accurate and highly reproducible.
Performance characterization of a 3D liquid scintillation detector for discrete spot scanning proton beam systems
Darne, Chinmay , Alsanea, Fahed , Robertson, Daniel , Sahoo, Narayan , Beddar, Sam , Robertson, Daniel G

Physics in Medicine & Biology | June 8, 2017 | Volume: 62 | Issue: 14 | Pages: 5652-5667

10.1088/1361-6560/aa780b Get It

Abstract Existing systems for proton beam dosimetry are limited in their ability to provide a complete, accurate, and detailed account of volumetric dose distribution. In this work, we describe the design and development of a portable, fast, and reusable liquid scintillator-based three-dimensional (3D) optical detection system for use in proton therapy. Our long-term goal is to use this system clinically for beam characterization, dosimetry, and quality assurance studies of discrete spot scanning proton beam systems. The system used a 20 × 20 × 20 cm 3 liquid scintillator volume. Three mutually orthogonal cameras surrounding this volume captured scintillation photons emitted in response to the proton beams. The cameras exhibited a mean spatial resolution of 0.21 mm over the complete detection volume and a temporal resolution of 11 ms. The system is shown to be capable of capturing all 94 beam energies delivered by a synchrotron and performing rapid beam range measurements with a mean accuracy of 0.073 ± 0.030 mm over all energies. The range measurement uncertainty for doses less than 1 cGy was found to be ±0.355 mm, indicating high precision for low dose detection. Finally, we demonstrated that using multiple cameras allowed for the precise locations of the delivered beams to be tracked in 3D. We conclude that this detector is capable of real-time and accurate tracking of dynamic spot beam deliveries in 3D. The high-resolution light profiles it generates will be useful for future 3D construction of dose maps.